基于Hausdorff距离的穴位阻抗重复性 被引量:1

Reproducibility of acupuncture point's impedance based on Hausdorff distance
下载PDF
导出
摘要 为了实现人体电阻抗的自动连续扫描,开发设计一套基于直角坐标机器人的穴位点与经络阻抗扫描系统.为了检测该阻抗扫描系统的稳定性与可重复性,提出采用Hausdorff距离来评价不同数据曲线的相似程度.选取人体背部两肩胛冈下角点之间的直线段作为扫描路径.采用同惠公司生产的TH2825阻抗分析器来测量阻抗数据.该实验中激励信号为幅值为1V、频率为1kHz的正弦波,量程设为100kΩ.直角坐标机器人经过调节可以沿着上述的直线段自动、连续地测量处于凝胶电极与银电极之间的阻抗.该实验共选取7名受试者,在25℃、湿度为50%的环境下进行测量.每名受试者在同一直线上进行5次连续扫描.采用Hausdorff距离来评估从同一名受试者获得的5次阻抗采样数据曲线的可重复性与相似性.实验结果表明,该阻抗测量系统具有良好的稳定性与可重复性. An impedance scanning system of acupuncture points and meridians based on Cartesian robot was designed in order to achieve the automatic body electrical impedance scanning. Hausdorff distance was applied to evaluate similarity between the different data curves in order to test the stability and reproducibility of the impedance scanning system. The same straight line was selected which lay in two scapular bottom points on human back as the scanning track. TH2825 LCR meter was used to measure impedance data produced in Tonghui company. The excitation signal was sine wave which amplitude was 1 V, frequency was 1 kHz and the measuring range was set at 100 kΩ. The Cartesian coordinate robot can be modulated to measure the impedance between the gel electrode and the Ag electrode along the abovementioned straight line automatically and continuously. Seven healthy subjects were selected in total, which were measured in the environment that temperature was 25 ℃ and moisture was 50℃. Each subject took five times of continuous scanning on the same straight line. The Hausdorff distance was applied in order to evaluate the reproducibility and similarity between the sampling impedance data curves which were acquired from the same subject. Experimental results showed that the reproducibility and stability of the impedance measurement system was favorable.
出处 《浙江大学学报(工学版)》 EI CAS CSCD 北大核心 2011年第10期1777-1780,1820,共5页 Journal of Zhejiang University:Engineering Science
关键词 HAUSDORFF距离 重复性 阻抗 直角坐标机器人 Hausdorff distance reproducibility impedance Cartesian-coordinate robot
  • 相关文献

参考文献10

  • 1AHN A C, COLBERT A P, ANDERSON B J. Electri- cal properties of acupuncture points and meridians: a systematic review [J]. Bioelectromagneties, 2008, 29(4) : 245 - 256.
  • 2COLBERT A P, YUN J, LARSEN A. Skin impedance measurements for acupuncture research: development of a continuous recording system[J]. Evidence-Based Com- plementary and Alternative Medicine, 2008, 5 ( 4 ).. 443 - 450.
  • 3MARTINSEN O G, GRIMNES S, MORKRID L. Line patterns in the mosaic electrical properties of human skin: a cross-correlation study[J]. IEEE Transactionson Biomedical Engineering, 2001, 48(6): 731 -734.
  • 4张国强,尹百顺.浅谈腧穴命名的特点[J].吉林中医药,2007,27(6):59-59. 被引量:7
  • 5段鲜红,郭现辉.浅谈背腰部穴位与背俞穴的不同应用[J].河南中医,2003,23(12):53-53. 被引量:7
  • 6崔承斌,王京京,吴中朝.从背俞穴与夹脊穴的关系论背俞功能带[J].中国针灸,2005,25(7):483-486. 被引量:197
  • 7吴新贵,何源浩.背俞穴主治作用的机理探讨[C]//中国针灸学会临床分会第14届全国针灸学术研讨会.Guil-in:中国针灸学会,2006:173-175.
  • 8苏妆 裴景春.背俞穴研究进展.中医药学刊,2005,.
  • 9维基百科:Hausdorffdistance[EB/OL].[2010-04-01].http:/en.wikipedia.orgwikiHausdorff—distance.
  • 10百度百科:Hausdorff距离[EB/OL].[2010-04-01].http://baike.baidu.comview2830255.html.

二级参考文献14

  • 1肖祖伟.针灸肺俞穴治疗风寒咳嗽40例[J].中国针灸,2003,23(11):667-667. 被引量:17
  • 2金镜,邵明宇,岳丹.针刺夹脊穴为主治疗下肢多发性神经炎32例[J].中国针灸,1997,17(3):138-138. 被引量:2
  • 3陈可冀.岳美中医学文集[M].北京:中国中医药出版社,2000.34-36,507,16-21.
  • 4孙维仁.背俞理论探索与临床应用[M].呼和浩特:内蒙古人民出版社,1994.3,4.
  • 5向之中.针灸经穴与原气[M].北京:中国中医药出版社,1994.46.
  • 6李定忠 李秀章.中医经络探秘[M].北京:解放军出版社,1999.59.
  • 7金百仁.五年来华佗夹脊穴临床应用研究进展[J].上海针灸杂志,1988,7(4):41-41.
  • 8刘俊红.肝俞穴点刺拔罐治疗顽固性麦粒肿32例[J].中国针灸,1999,19(11):680-680.
  • 9李常度.华佗夹脊穴的作用机理及临床应用[J].贵阳中医学院学报,1988,(4):23-23.
  • 10邵水金 谢强.魏稼针灸经验集[M].上海:上海中医药大学出版社,2000.50.

共引文献206

同被引文献14

引证文献1

二级引证文献26

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部