摘要
A method for simultaneous analysis of five tetracyclines(tetracycline,oxytetracycline,chlortetracycline,doxycycline and minocycline) in bovine muscle was developed by UPLC-MS/MS.The sample was extracted with Mcllvaine buffer-EDTA(pH 4.0) and cleaned up by solid-phase extraction on Oasis HLB cartridges.Analytes were separated by UPLC,and detected by mass spectrum with MRM mode.The calibration curves showed good linear relations in the concentration range of 1-100 μg·L-1(r>0.994) for five tetracyclines.The average recoveries at the spiked concentration levels of 1,20,50 and 100 μg·kg-1 were range of 68%-105% and the RSD were less than 34.1%.The LOD of the method was 0.05 μg·kg-1 for five tetracyclines.
A method for simultaneous analysis of five tetracyclines(tetracycline,oxytetracycline,chlortetracycline,doxycycline and minocycline) in bovine muscle was developed by UPLC-MS/MS.The sample was extracted with Mcllvaine buffer-EDTA(pH 4.0) and cleaned up by solid-phase extraction on Oasis HLB cartridges.Analytes were separated by UPLC,and detected by mass spectrum with MRM mode.The calibration curves showed good linear relations in the concentration range of 1-100 μg·L-1(r>0.994) for five tetracyclines.The average recoveries at the spiked concentration levels of 1,20,50 and 100 μg·kg-1 were range of 68%-105% and the RSD were less than 34.1%.The LOD of the method was 0.05 μg·kg-1 for five tetracyclines.
出处
《分析测试学报》
CAS
CSCD
北大核心
2008年第S1期197-199,共3页
Journal of Instrumental Analysis