单根电极导管VDD起搏系统的初步临床应用

InitialClinicalExperienceWithaSingleleadVDDPacingSystem/HuangWenxin,YangYuezhong,ChenJianzhong,etal.DepartmentofCardiology,AfiliatedHospitalofGuilinMedicalColege,Guilin,541001AbstractAsingleleadVDDpacingsystem(DromosSLandSL60/13UP)wasimplantedin4patientsw
全文增补中
导出
摘要 对4例窦房结正常的患者(完全性房室传导阻滞3例,肥厚梗阻性心肌病1例)植入单根电极导管VDD起搏系统,并进行随访观察,结果显示:术中所测平均A波振幅为2.3±1.3mV,随访期间(8~12个月,平均10.5±1.7个月),房室同步起搏率保持在100%,无心房感知不良、肌电干扰和起搏器介导性心动过速等并发症。表明单电极导管VDD起搏系统具有植入简单和房室同步功能稳定的特点。 A singlelead VDD pacing system (Dromos SL and SL 60/13UP)was implanted in 4 patients with normal sinoatrial node function (3 with complete AV block and 1 with IHSS). The results showed that the mean amplitude of P wave was 2.3±1.3mV at implantation, during the followup(10.5±1.7 months on average),there was no significant chang in atrial sensing threshold(P>0.05) and the percentage of atrioventricular synchronization was maintained at 100%. Postoperative complications such as atrial undersensing,myopotential interference and PMT were not observed. According to data in this literature, a single lead VDD pacing system is an optimal treatment for patient with high degree AVB and normal sinoatrial node function.
出处 《华夏医学》 1998年第5期12-13,共2页 Acta Medicinae Sinica
关键词 单电极VDD起搏器 心脏传导阻滞 心房同步起搏 singlelead VDD pacing heart block atrial triggered ventricular pacing