期刊文献+

Determination of CaO in Baotou Columbite and Steel Cinder with Flame Atomic Absorption Spectrometry

Determination of CaO in Baotou Columbite and Steel Cinder with Flame Atomic Absorption Spectrometry
下载PDF
导出
摘要 The determination of CaO content in columbite and steel cinder with flame atomic absorption spectrometry is studied. EDTA+TEA is used to eliminate the interferences, in HCI media,with La as releaser. The methods of sample treatment and the CaO in remainder undissolved in acids have been conducted. The result of the determination and recovery of CaO shows that the rate of recovery is 100% ~ 102 %, R. S.D<2 %. The determination of CaO content in columbite and steel cinder with flame atomic absorption spectrometry is studied. EDTA+TEA is used to eliminate the interferences, in HCI media,with La as releaser. The methods of sample treatment and the CaO in remainder undissolved in acids have been conducted. The result of the determination and recovery of CaO shows that the rate of recovery is 100% ~ 102 %, R. S.D<2 %.
出处 《International Journal of Minerals,Metallurgy and Materials》 SCIE EI CAS CSCD 1997年第2期62-62,共1页 矿物冶金与材料学报(英文版)
关键词 atomic absorption spectrometry CaO columbite steel cinder atomic absorption spectrometry, CaO,columbite, steel cinder