输出因子的两种测量法比较 被引量:2

Comparison between two methods for the measurement of the output factor
原文传递
导出
摘要 目的研究固定机器量和固定信号采集时间两种不同条件下测得的输出因子差别。材料与方法在VarianClinic-6和Mevatron77电子直线加速器下采用固定机器输出量和固定测量电离室信号采集时间两种方法测量了输出因子。结果两种结果间最大可有1.2~1.4%的差别,且随射野增大而减少。结论根据机器结构和射线与物质相互作用原理分析,产生这种差异的原因是在光栏上表面的反散射在机器电离室内产生的电离所致。 Purpose To evaluate the difference between the fixed monitor unit and fixed signal collecting time methods in measuring of output factors. Materials and Methods With a fixed monitor unit or a fixed signal collecting time,the output factors were measured under a Varian clinic-6 and a Mevatron 77 linear accelerators. Results The maximum difference between these two methods was 1.2~1.4%which had the decreasing tendency when the radiation field increased. Conclusion This phenomena is due to the back-scatter particles penetration into the beam monitor chamber from the upper surfaces of the collimator and causes its counting according to the structure of the machine and the principle of the interaction between photon and material.
出处 《中华放射肿瘤学杂志》 CSCD 1996年第4期261-263,共3页 Chinese Journal of Radiation Oncology
关键词 机器电离室 反散射 输出因子 Bearn monitor chamber Back-scatter Output factor
  • 相关文献

同被引文献9

  • 1李黎军,朱海军,张新忠,李飞舟,宋宏羽.剂量计算公式总散射校正因子测量深度的研究[J].中华放射肿瘤学杂志,2004,13(3):238-240. 被引量:2
  • 2戴建荣.X(γ)射线射野剂量学.见:胡逸民主编.肿瘤放射物理学[M].北京:原子能出版社,1999.198-205.
  • 3乔林,杨志刚,刘文杰等.Visual C++6.0程序设计[M].北京:中国铁道出版社.2001.87-164.
  • 4Day MJ.The equivalent field method of axial dose determi nations in rectangular fields,in Central Axis Depth Dose Data in Radiotherapy[J].Br J Radiol,1972(Suppl),11-95.
  • 5IEC.IEC 1217.Radiotherapy equipment:coordinates,movements and scales[M].1996.5-8.
  • 6Day MJ.A note on the calculation of dose on X-ray field[J].Br J Radiol,1950,23:270-275.
  • 7Biggs PJ,Boyer AL,Doppke KP.Electron dosimetry of irregular fields on the Clinac-18.Int[J].J Radiat Oncol Phys,1979,5:433-435.
  • 8Sterling TD,Perry H,Katz I.Derivation of a mathematical expression for the percent depth dose of cobalt 60 beams and visualization of multiple field dose distributions[J].Br J Radiol,1964,37:544-548.
  • 9韩树奎,路长春.直线加速器上、下铅门对准直器散射因子的影响[J].中华放射肿瘤学杂志,1996,5(3):63-64. 被引量:4

引证文献2

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部