X(γ)刀剂量场的测量 被引量:4

Dosimetric measurements for X(γ)-knife beams
原文传递
导出
摘要 目的 :目前 ,国内X、γ刀的输出剂量及剂量分布尚无适宜的测量设备及相应的技术规范。由于X、γ刀剂量场的特点 ,不能用常规的电离室测量。本文的目的是为X、γ刀的剂量测量提供合理的设备和方法。材料与方法 :研制了灵敏区为 1mm的半导体探测器和体积为 0 .0 3cc的石墨电离室。用该探测器测量了γ刀在各种模体中的输出剂量及剂量分布。结果 :用研制的SCD -61半导体剂量仪和 0 .0 3cc石墨电离室测量了不同模体、不同准直器的输出剂量及剂量分布 ,根据对球形模的测量值 ,计算出各种射野时头形模的量分布 ,用热释光剂量仪做了验证 ,焦点值均在 5 %范围内一致。结论 :X、γ刀的输出剂量及剂量分布应在现场进行实际测量 ,所用探头直径应小于射野直径的一半。灵敏区~ 1mm的半导体探测器适合作X。 Purpose: There is no appropriate instrument or devices and corresponding technological specification for measurement of output dose and dose disdribution in stereotactic radiothrapy. Small field of the stereotatic radiotherapy can't be measured by normal ionization chamber for its characteristics.A practible instrument and method were developed for dose measurement in stereotactic irradiation.Materials and Methods:A diode detector with 1mm sensitivety volume and a 0.3cc graphite chamber developed by us are used to measure the dose output & dose distribution in warious phantom for the beams used in stereotactic irradiation.Results:The dose outputs and dose ditributions measured by SCD-61 diode developed by us were tested and verified by TLD dosemeter,and the results show tha there is a 5% consistency for the value at the focus point.Conclusion: the diameter of the detector used for measurement must be smaller than the halt size of the radiation beams to be measured. A diode detector with 1mm sensitivity volume is the best choice for the dosimetric measurements for smaller beams used in stereotactic irradiation.
出处 《中华放射肿瘤学杂志》 CSCD 1996年第2期27-31,共5页 Chinese Journal of Radiation Oncology
关键词 X(γ)刀剂量场 剂量分布 半导体探测器 X(γ)-knife beams Dose distributions Diode detector
  • 相关文献

同被引文献13

  • 1张绍刚,刘明远.胶片法在测量中的几个问题[J].中国医学物理学杂志,1996,13(3):160-162. 被引量:5
  • 2彭振军.医用磁共振成像技术[M].武汉:湖北科学技术出版社,1997..
  • 3王所亭 林光方 等.X刀剂量学数据的测量与研究.301医院内部资料[M].,..
  • 4杨开军,颅脑疾病X刀治疗学,1997年,149页
  • 5彭振军,医用磁共振成像技术,1997年,130页
  • 6王肖兰,中华放射肿瘤学杂志,1996年,5期,97页
  • 7Uesr's manual of Leksell Gamma Unit[M],Elekta Edition,April 1992.
  • 8Leksell L.Stereotactic radiosurgery[J].J Nourol NeurosurgPsych,1983, 46:797.
  • 9Robert J.et al. The application accuracy of steroereotactic frames[J].Neurosurgery. 1994,35:682.
  • 10Yeung D. Palta J. Fontanesi J. et al. Systematic analysis of errors in targetlocalization and treatment delivery in stereotactic radiosurgery (SRS)[J]. Int JRadiat Oncol Biol Phys, 1993,28:493

引证文献4

二级引证文献10

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部