Perkins眼压计与Goldmann及Schiotz式眼压计之对比研究 被引量:1

COMPARISON BETWEEN PERKINS TONOMETER AND COLDMANN AND SCHIOTZ TONOMETERS
下载PDF
导出
摘要 Goldmann压平眼压计是目前国际上公认的最精确的眼压计。国内外文献已证明了Perkins眼压计具有同Goldmann眼压计同样的准确性,本文通过对正常人35例70只眼的Goldmann眼压,Perkins眼压(坐卧位)及Schiotz眼压测定之比较,证实了Goldmann眼压与Perkins眼压计同样准确(统计学处理P>0.05)。Perkins眼压与Schiotz眼压之测定值及校正值相比。统计学上均具有显著性差异。说明Schiotz眼压计在测定眼压时,具有无法克服之巩膜硬度和角膜屈率之影响而远不如Perkins眼压计准确。但二者之相关关系呈正相,说明Schiotz眼压计也可粗略估计眼压。应尽可能用Perkins眼压计取替Schiotz眼压计。 It is well known that Goldmann tonometer is the most precise by now. Perkins tonometer is as precise as.Goldmarn, which has been proved . It had been .examinat-ed that the optical pressure in 70 eyes of 35 peoples detected with Golclmarp, Perkins and Schiotz tonometers. It was observed that there was no significant difference between the values of the optical pressure detected with Goldmann and Perkins tonometer (P>0, 05). Compared with the values of Perkins optical pressure, the measured and rectified values of Schiotz optical pressure were very different, which expressed that Schiotz tonometer was less accurate than Perkins and it could not overcome the effects of the sclera hardness and cornea . flexure But the relationship between the values of optical pressure detected with Perkins tonometer and Schiotz one was positive correlation. Therefore, Schiotz tonometer can be used to measure roughly optical pressure; It is suggested that Schiotz tonometer should be replaced by Perkins tonometer as far as possible.
出处 《西南国防医药》 CAS 1994年第1期13-14,63-64,共4页 Medical Journal of National Defending Forces in Southwest China
关键词 Perkins眼压计 GOLDMANN眼压计 SCHIOTZ眼压计 健康人眼压 Perkins tonometer, Goldmarn tonometer,Schiotz tonometer, optical pressure
  • 相关文献

参考文献1

  • 1水迎波,魏厚仁.体位变化对正常眼眼压的影响[J]眼科研究,1987(03).

同被引文献12

  • 1Quigley HA, Broman AT. The number of people with glaucoma worldwide in 2010 and 2020[J]. Br J Ophthalmol, 2006.90(3): 262-267.
  • 2L Liang YB, Friedman DS, Zhou Q, Yang X, et al. Prevalence of primary open angle glaucoma in a rural adult Chinese population: the Handan eye study[J]. Invest Ophthalmol Vis Sci, 2011, 52(11): 8250-8257. DOI: 10. 1167/Joys. 11-7472.
  • 3Sultan MB, Mansberger SL, Lee PP. Understanding the importance of IOP variables in glaucoma: a systematic review[J]. Surv Ophthalmol, 2009, 54(6): 643-662. DOI: 10. 1016/ j. survophthal. 2009.05. 001.
  • 4Bengtsson B. comparison of sehiotz and Goldmann tonometry in a population[J]. Aeta 0phlhalmol (Copenh), 1972, 50(4): 445-457.
  • 5Grewal DS , Stinnett SS , Folgar FA, et al. A tomparatwe Study of Rebound Tonometry With Tonopen and Goldmann Applanation Tonometry Following Vitreoretinal Surgery[J]. Am J Ophthalmol, 2016, 161: 22-28. DOI: 10. 1016/j. ajo. 2015.09. 022.
  • 6Bhartiya S, Bali SJ, James M, et al. Test retest variability of TonoPen AVIA[J]. Indian J Ophthalmol, 2013, 61(3): 129- 131. DOI: 10. 4103/0301-4738. 109384.
  • 7Weinreb RN, Liu JH. Nocturnal rhythms of intraoeular pressure[J]. Arch Ophthalm01, 2006, 124(2): 269-270.
  • 8Colombo L, Fogagnolo P, Montesauo G, et al. , Strategies to estimate the characteristics of 24-hour IOP curves of treated glaucoma patients during office hours[J]. BMC Ophthalmol, 2016, 16(1): 15. DOI: 10. 1186/s12886-016-0191-7.
  • 9范芸,潘英姿,朱赛楠,李梅,乔荣华,才瑜,方圆,刘丽娜,王捷.iCare回弹式眼压计的临床应用评价[J].眼科研究,2010,28(11):1074-1077. 被引量:7
  • 10何跃,陈洁,吕红彬,张曙光,李艳梅,袁援生.iCare回弹式眼压计与Goldmann压平眼压计测量结果的比较[J].眼科研究,2010,28(12):1162-1165. 被引量:5

引证文献1

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部