期刊文献+

CARBON WIRE AMPEROMETRIC FLOW THROUGH DETECTOR

CARBON WIRE AMPEROMETRIC FLOW THROUGH DETECTOR
下载PDF
导出
摘要 A carbon wire amperometric flow through detector is described which is easily constructed at low cost for application to flow injection analysis,continuous stream analysis and liquid chromatography.The theoretical limiting current of the detector is 2.14×10~6nCD^(2/3)u_f^(1/3) with 36% actual current efficiency.By use of flow injection analysis,the detecting limits for Ag(I)and Cr(VI)had achieved 1 ppb and 0.4 ppb respectively with 1.2%(n=20)relative standard deviation after preliminary experiments. A carbon wire amperometric flow through detector is described which is easily constructed at low cost for application to flow injection analysis,continuous stream analysis and liquid chromatography.The theoretical limiting current of the detector is 2.14×10~6nCD^(2/3)u_f^(1/3) with 36% actual current efficiency.By use of flow injection analysis,the detecting limits for Ag(I)and Cr(VI)had achieved 1 ppb and 0.4 ppb respectively with 1.2%(n=20)relative standard deviation after preliminary experiments.
出处 《Chinese Chemical Letters》 SCIE CAS CSCD 1990年第2期113-116,共4页 中国化学快报(英文版)
基金 This Project is supported by the National Science Foundation of China