透析用水中游离氯与氯胺的测定 被引量:4

Detection of chlorine and chloramines in dialysis water
下载PDF
导出
摘要 目的透析用水中游离氯和氯胺浓度超标会对人体造成巨大危害,而我国目前大多数透析中心只能检测游离氯,本研究通过检测自来水和透析用水中游离氯和氯胺含量,评估了单独检测游离氯的安全性,并评估了目前常用的TMB目视比色法的准确性。方法以DPD分光光度法为金标准,同时采用DPD分光光度法和TMB目视比色法对北京市9家透析中心的自来水和透析用水中游离氯和氯胺浓度进行检测,每家血液透析中心检测2d,每天检测10次,每次间隔1h;对透析用水中游离氯和氯胺浓度进行相关性分析,P<0.05认为有统计学意义;对DPD分光光度法和TMB目视比色法的检测结果进行Kappa检验。结果在各个时点,各透析中心的透析用水中游离氯和氯胺浓度均在国家标准范围之内;透析用水中游离氯和氯胺浓度之间不存在线性相关关系(r=0.14,P=0.146);同DPD分光光度法相比,TMB目视比色法所测得游离氯结果的假阴性率为100%,kappa值为0,所测总氯结果的假阴性率10.67%,kappa值为0.88。结论单独检测游离氯是不安全的,必须同时检测总氯浓度;同DPD分光光度法相比,TMB目视比色法测得的游离氯和总氯结果存在一定的假阴性率,建议有条件的单位尽量采用DPD分光光度法。 Objective Exceeding chlorine and chloramines in dialysis water is harmful to human body, but most dialysis centers in China can only measure chlorine content. The aim of the current study was to evaluate the safety only based on chlorine content, as well as the accuracy of TMB visual colorimetry. Methods Levels of chlorine and chloramines in dialysis water and tap water were measured using both TMB visual colorimetry and DPD spectrophotometry, and values from DPD spectrophotometric method were used as the golden standards. Dialysis water and tap water from 9 dialysis centers were assayed every hour and ten times per day for two days. Linear regression was used to explore the relationship between chlorine and chloramines in dialysis water, and P 0.05 was considered statistically significant. Kappa test was used to compare the difference between the values from DPD spectrophotometry and those from TMB visual colorimetry. Results Concentrations of chlorine and chloramines in dialysis water were below the standards recommended by AAMI at every time point. There was no linear relationship between chlorine and chloramines contents in dialysis water (r =0.14, P =0.1460). TMB visual colorimetry gave a false negative rate of 100% with kappa=0 for free chlorine measurement, and a false negative rate of 10.67% with kappa=0.88 for total chlorine measurement as compared with the values from DPD spectrophotometry, Conclusion For dialysis water, monitoring free chlorine alone is probably unsafe, and chloramines or total chlorine should also be measured. Compared to DPD spectrophotometry, TMB visual colorimetry gives a remarkable false negative rate, suggesting the preferential use of DPD spectrophotometry method if possible.
出处 《中国血液净化》 2011年第11期588-593,共6页 Chinese Journal of Blood Purification
关键词 游离氯 氯胺 透析用水 Chlorine Chloramines Dialysis water
  • 相关文献

参考文献18

  • 1左力.透析用水和患者安全[J].中国血液净化,2009,8(1):1-4. 被引量:18
  • 2顾军农,张春雷,王丹,罗亮,杨胜武.预处理技术在北京自来水厂的应用与分析[J].中国给水排水,2006,22(z1):202-206. 被引量:3
  • 3Thomas W. Stief. The physiology and pharmacology of singlet oxygen[J]. Med Hypotheses, 2003, 60:56?-5?2.
  • 4Junglee NA, Rahman SU, Wild M, et al. When pure is not so pure:ehloramine-related hemolytic anemia in home hemodi alysis patients[J].Hemodial Int, 2010,14:327-332.
  • 5Tipple MA, Shusterman N, Bland LA, et al. Illness in hemodialysis paitents after exposured to ehloramines contaminated dialysate[J]. ASAIO Trans, 1991, 3?:588 -591.
  • 6de Oliveira RM, de los Santos CA, Aatonello l, et al. Warning: an anemia outbreak due to ehloramines exposure in a clean hemodialysis unit an issue to be revisited[J]. Ren Fail, 2009,31:81-83.
  • 7Fluck S, McKane W, Cairns T, et al. Chloramine induced haemolysis presenting as erythropoietin resistance[J]. Nephrol Dial Transplant, 1999,14:1687-1691.
  • 8Calderaro RV, Heller L. Outbreak of hemolytic reactions associated with chlorine and chloramines residuals in hemodialysis water[J]. Rev Saude Publica, 2001,35:481- 486.
  • 9朱彩凤,陈悦霞,张红梅.透析用水氯胺超标所致不良反应32例临床分析[J].浙江医学,1998,20(1):43-44. 被引量:4
  • 10Association for the Advancement of Medical Instrumentation [M]. 1982. AAMI standards and Recommended Practices, 2008 Ed. AAMI, Arlington, VA, 2008.

二级参考文献17

  • 1李保春,翟振燕.上海市血液净化中心透析用水及透析液细菌污染情况调查[J].中国血液净化,2005,4(1):52-54. 被引量:23
  • 2[1]刘文君.微污染源水饮用水处理[M].北京:中国建筑工业出版社,1999.
  • 3Amato RL.Water treatment for hemodialysis-updated to include the latest AAMI standards for dialysate (RD 52:2004).Nephrol Nur J,2005,32:151-169
  • 4Cappelli G,Ravera F,Ricardi M,et al.Water treatment for hemodialysis:A 2005 update.Contrib Nephrol,2005,149:42-50
  • 5Pontoriero G,Pozzoni P,Andrulli S,et al.The quality of dialysis water.Nephrol Dial Transplant,2003,18(Suppl 7):21-25
  • 6Pizzarelli F,Cerrai T,Biagini M,et al.Dialysis water treatment systems and monitoring in Italy:Results of a national survey.J Nephrol,2004,17:565-569
  • 7北京市卫生局和北京市血液透析质量控制和改进中心.北京市血液透析质量管理规范,2003,1-28
  • 8Burwen,DR,Olsen,SM,Bland,LA,et al.Epidemic aluminum intoxication in hemodialysis patients traced to use of an aluminum pump.Kidney Int,1995,48:469
  • 9Klein E,Pass T,Harding GB,et al.Microbial and endotoxin contamination in water and dialysate in the central United States.Artif Organs,1990,14:85-94
  • 10Laurence RA,Lapierre ST.Quality of hemodialysis water:a 7-year multicenter study.Am J Kidney Dis,1995,25:738-750

共引文献48

同被引文献19

  • 1左力.透析用水和患者安全[J].中国血液净化,2009,8(1):1-4. 被引量:18
  • 2王质刚.血液净化学[M].3 版.北京:科学技术出版社,2010:857-867.
  • 3刘萌,吕晓,刘慧松.品管圈模式对血液透析患者实施健康教育的效果观察[J].实用心脑肺血管病杂志,2013,21(2):153.
  • 4高翔,梅长林.水处理系统及透析液//梅长林,叶朝阳.实用透析手册[M].北京:人民卫生出版社,2003:29-37.
  • 5季大玺.血液透析原理与设备//黎磊石,刘志红.中国肾脏病学[M].北京:人民军医出版社,2008:1445-1458.
  • 6游志清,徐永堂.从AMMl2005谈血液透析用液规范[J].血液净化医学杂志,2006,11(1):45-61.
  • 7Bernad M, Charles MM. Water treatment for contemporary emodiaslysis//Winchester JF, eds. Replacement of renal function by dialysis[M]. 4th ed. Dorderchet: Kluwer Academic Publisher, 1996 : 231-255.
  • 8Lonnemann G, Koch KM. Efficacy of ulter pure dialysate in the therapy and prevention of Haemodialysis-associated amyloidosis[J]. Nephrol Dial Transplant, 2001,16 ( Suppl 4 ) : 17-22.
  • 9Canaud B, Bosc JY, Leray H, et al. Microbiologic purity of dialysate: rationale and technical aspects[J]. Blood Purif, 2000, 18(3) :200-213.
  • 10Vaslaki L,Karatson A,Voros P,et al. Can sterile and pyrogen free on-line substitution fluid be routinely delivered? A muhicentric study on the microbiological safety of on-line heamodiafiltration [ J ]. Ncphrol Dial Transplant, 2000, 15 (Suppl 1) :74-78.

引证文献4

二级引证文献8

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部