采用欧洲腰椎体模对多中心腰椎定量CT的精密度和准确度评估 被引量:13

Assessment of Precision and Accuracy of Spinal QCT in a Multi-centre Clinical Research with a European Spine Phantom
下载PDF
导出
摘要 目的采用欧洲腰椎体模(ESP)作为标准评估多中心腰椎定量CT(QCT)骨密度测量的重复性和准确性误差。材料与方法2个中心(3台CT分别是东芝64排CT、东芝16排CT及GE64排CT)按照常规腰椎定量CT骨密度测量条件和方法对欧洲腰椎体模(ESP)进行重复测量各10次;另外4个中心(CT分别是东芝64排、东芝16排、西门子64排、GE双排)分别用欧洲体模(ESP)扫描测量一次。结果 3台QCT(CT顺序依次是东芝64排,东芝16排,GE64排)对欧洲体模测得的相对精密度CV分别是0.512%、0.523%、0.514%。7台QCT机器间的精密度误差CV分别是L14.213%、L22.752%、L34.786%。7台CT的准确度变化范围和平均值(括号因数值)分别为L11.20%~8.79%(5.1);L2-0.25%~9.15%(3.8);L3-7.03%~5.69%(-0.5)。结论参与研究的多中心QCT系统的机器内精密度、机器间精密度及准确度较好,结果准确可靠。 Purpose To evaluate the reproducibility and accuracy of spinal QCT in a multi-centre clinical research using a European Spine Phantom. Material and Methods ESP was scanned 10 times using routine technique without repositioning the phantom on 3 CT scanners (64-row Toshiba, 16-row Toshiba and 64-row GE CT scanner) in two centers. The ESP was scanned once following the same protocols in another four centers using 64-row Toshiba, 16-row Toshiba, 64-row Siemens and 2-row GE CT scanner respectively. Results The coefficient of variation of the first three QCT systems was 0.512%, 0.523% and 0.514% respectively for integral phantom. The inter-modality precision errors of seven QCT systems were 4.213%, 2.275%, and 4.786% in L1, L2 and L3 respectively. The range (and mean) of accuracy error of seven QCT systems were respectively 1.20% to 8.79% (5.1), -0.25% to 9.15% (3.8) and -7.03% to 5.69% (-0.5) in L1, 1-2 and L3. Conclusion Spinal QCT shows excellent precision, intermodality precision and accuracy in this multi-centre clinical research.
出处 《中国医学影像学杂志》 CSCD 北大核心 2011年第12期912-917,共6页 Chinese Journal of Medical Imaging
基金 北京市卫生系统高层次卫生技术人才培养项目资助(2009-02-03)
关键词 骨密度 腰椎 体层摄影术 X 线计算机 欧洲腰椎体模 Bone density Lumbar vertebrae Tomography, X-ray computed European spine phantom
  • 相关文献

参考文献25

  • 1Osteoporosis: review of the evidence for prevention, diagnosis and treatment and cost-effective analysis. Osteoporos Int, 1998, 8(Suppl4): S3-S6.
  • 2Adams JE. Quantitative computed tomography. Eur J Radiol, 2009, 71(3): 415-424.
  • 3Zhu H, Fang J, Luo X, et al. A survey of bone mineral density of healthy Han adults in China. Osteoporos Int, 2010, 21(5): 765-772.
  • 4Assessment of fracture risk and its application to screening for postmenopausal osteoporosis. Report of a WHO Study Group. World Health Organ Tech Rep Ser, 1994, 843: 1-129.
  • 5Bonjour JP, Ammann P, Rizzoli R. Importance ofpreclinical studies in the development of drugs for treatment of osteoporosis: a review related to the 1998 WHO guidelines.Osteoporos Int, 1999, 9(5): 379-393.
  • 6Prevention and management of osteoporosis. World Health Organ Tech Rep Ser, 2003.1-164.
  • 7Lees B, Garland SW, Walton C, et al. Evaluation of the european spine phantom in a multi-centre clinical trial. Osteoporos Int, 1997, 7(6): 570-574.
  • 8Gluer CC, Blake G, Lu Y, et al. Accurate assessment of precision errors: how to measure the reproducibility of bone densitometry techniques. Osteoporosis Int, 1995, 5(4):262-270.
  • 9Young M. Cross-calibration of axial DEXA systems. Radiography, 2001, 7(1): 79-83.
  • 10Tothill P, Hannan WJ. Precision and accuracy of measuring changes in bone mineral density by dual-energy X-ray absorptiometry. Osteoporos Int, 2007, 18(11): 1515-1523.

二级参考文献37

  • 1吴胜勇,张美超,李景学,欧阳钧,温连庆,祁吉.骨质疏松老年妇女有限元模型的建立[J].中国临床解剖学杂志,2004,22(6):661-663. 被引量:8
  • 2吴胜勇,杨立,祁吉,王斌,温连庆,李景学.骨质疏松老年妇女腰椎骨密度及结构的多层螺旋CT研究[J].中华放射学杂志,2005,39(11):1165-1170. 被引量:28
  • 3秦明伟,余卫,孟迅吾,邢小平,徐苓,田均平,曹坚,严洪珍,葛秦生.正常人腰椎骨密度变化──445例QCT测量分析[J].中国医学科学院学报,1996,18(6):439-443. 被引量:19
  • 4Ruegsegger P, Elsasser U, Anliker M, et al. Quantification of bone mineralisation using computed tomography. Radiology, 1976, 121:93-97.
  • 5Genant HK, Cann CE, Ettinger B, et al. Quantitative computed tomography of vertebral spongiosa : a sensitive method for detecting early bone loss after oophorectomy. Ann Intern Med, 1982, 97: 699 -705.
  • 6Cann CE, Genant HK. Precise measurement of vertebral mineral content using computed tomography. J Comput Assist Tomogr, 1980, 4:493-500.
  • 7Cann CE. Quantitative CT applications: comparison of current scanners. Radiology, 1987, 162(1 Pt 1):257-261.
  • 8Boden SD, Goodenough DJ, Stockham CD, et al. Precise measurement of vertebral bone density using computed tomography without the use of an external reference phantom. J Digit Imaging, 1989, 2:31-38.
  • 9Gudmundsdottir H, Jonsdottir B, Kristinsson S, et al. Vertebral bone density in Iceland women using quantitative computed tomography without an external reference phantom. Osteoporos, 1993,3:84-89.
  • 10Hopper KD, Wang MP, Kunselman AR. The use of clinical CT for baseline bone density assessment. J Comput Assist Tomogr, 2000, 24:896-899.

共引文献17

同被引文献87

引证文献13

二级引证文献1293

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部