反射式额头血氧饱和度测量及其在军事医学中的应用 被引量:1

Oxygen Saturation Measurement in Forehead Using Reflection Mode in Military Medicine
下载PDF
导出
摘要 目的:设计一种反射式额头血氧饱和度测量装置,研究其在军事医学中的应用。方法:根据反射式血氧饱和度测量原理提出反射式血氧饱和度传感器设计方案,并分析血氧饱和度传统计算方法的误差及消除方法。结果:该血氧饱和度测量装置测量精度:在70%~100%范围内无运动时为±2%,有运动时为±3%。该装置已经在多个军事课题中得到应用。结论:反射式额头血氧饱和度测量技术对在军事作业环境下了解人体生命体征和评定功能状态具有重要意义。 Objective To design an oxygen saturation measurement apparatus in the forehead using reflection mode and explore its application in military medicine.Methods Both hardware design schemes for oxygen saturation measurement apparatus in the forehead using reflection mode and error analysis in traditional methods for calculating oxygen saturation,in which there was a novel method for error correction,were presented.Results The oximeter's accuracy was within ±2% when without movement,±3% when moving in the range of 70%~100%.The oximter had been used in multiple military tasks.Conclusion Oxygen saturation measurement in the forehead using reflection mode can be used to find out and to assess the soldier's physiological status in military tasks.
作者 吕沙里 李晔
出处 《医疗卫生装备》 CAS 2011年第11期27-30,共4页 Chinese Medical Equipment Journal
关键词 血氧饱和度 光电容积脉搏波描记(PPG) 额头 透射式 反射式 oxygen saturation photoplethysmographic(PPG) forehead transmission mode reflection mode
  • 相关文献

参考文献8

  • 1PUJARY C J. Investigation of photodetector optimization in reducing power consumption by a noninvasive pulse oximeter sensor[D/OL]. WORCESTER POLYTECHNIC INSTITUTE,2004. [2011-03 -15]. https ://www.wpi.eduPubsETDAvailableetd -0116104 - 145848unrestrictedcpujary.pdf.
  • 2Schmitt J M. Simple Photon Diffusion Analysis of the Effects of Multiple Scattering on Pluse Oximetry [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1991,38 (12) : 1 194-1 203.
  • 3MANNHEIMER P D,CASCIANI J R,FEIN M E. Wavelength Selection for Low-Saturation Pluse Oxymetry[J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1997,44(3) : 148-158.
  • 4邹波.利用动脉血液的时变差分光谱实现血氧饱和度的测量[J].红外,2009,30(8):44-48. 被引量:2
  • 5宋俊杰,王哲,金海龙,王超.三波长低血氧测量系统的研究与实现[J].激光与红外,2010,40(1):62-65. 被引量:4
  • 6王建军,孙万蓉,张旭.通过频谱分析测量血氧饱和度[J].生物医学工程研究,2004,23(3):167-169. 被引量:4
  • 7TEXAS INSTRUMENTS INCORPORATION, MSP430x4xx Family User' s Guide SLAU056AF[EB/OL]. (2006-03-15) [2011-03-20]. http://www.ti.com/msp430.
  • 8SHENDERB S,FORSTERE M,HREBIEN L,etal.Acceleration-Induced Near-Loss of Consciousness:The "A-LOC"Sysdrome, Aviation, Space, and Environmental Medicine[J]. 2003,74( 10):1021-1028.

二级参考文献13

共引文献7

同被引文献12

引证文献1

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部