C-L短柱瓣体外流体动力学评价 被引量:9

An assessment of hydrodynamics of the C-L pugestrut valve prosthesis in vitro
下载PDF
导出
摘要 目的:评价C-L短柱瓣体外流体动力学性能。方法:以C-L标准瓣、Medtronic-Hall瓣作对照,对C-L短柱瓣进行了静态泄漏、定常流及脉动流测试。结果:C-L短柱瓣静态泄漏率为(298±66)ml/min,符合临床应用要求。C-L短柱瓣定常流跨瓣压差明显低于C-L标准瓣(P<0.01),流场模型显示其瓣后流场涡流明显小于C-L标准瓣,且大瓣口层流清晰,小瓣口基本保持层流。脉动流测试表明C-L短柱瓣与Medtronic-Hall瓣跨压差相近,每搏返流量相近。结论:C-L短柱瓣体外流体动力学性能优良,明显优于第一代C-L侧倾碟瓣,并与国外同类产品相近,是一种有希望的新一代国产人工心脏瓣膜。 Objective: To assess the hydrodynamic performance of C L pugestrut valve prosthesis in vitro . Methods: The leakage, steady flow and pulsatile flow of the valve prosthesis was tested in vitro with the C L standard valve and Medtronic Hall valve prosthesis as control. Results: The rate of leakage of the C L pugestrut valve was (298±66) ml/min, accorded with the demand of clinical use. The steady flow pressure drop across the C L pugestrut valve was significantly lower than the C L standard valve( P <0 01). The flow pattern showed that the whirling flow behind the C L pugestrut valve was much gentler than the C L standard valve, with clear laminated flow through the big orifice and almost liminated flow through the small orifice. The pressure dropped and stroke recurrent volume of the C L pugestrut valve on pulsatile flow test was similar to Medtronic Hall valve. Conclusion: The hydrodynamic performance of the C L pugestrut valve in vitro is excellent, superior to the C L standard valve, similar to Medtronic Hall valve. This heart valve prosthesis would be an new potenitial model of the mechanical valve prostheses made in China.
出处 《第二军医大学学报》 CAS CSCD 北大核心 1999年第12期1006-1008,共3页 Academic Journal of Second Military Medical University
基金 全军"九五"医药卫生科研基金!项目编号96 Z0 3 1
关键词 C-L短柱瓣 流体动力学 人工心脏瓣膜 C L pugestrut hydrodynamics heart valve prosthesis
  • 相关文献

参考文献1

  • 1邹良建,上海生物医学工程,1993年,43卷,3期,6页

同被引文献86

引证文献9

二级引证文献21

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部