期刊文献+

辐射剂量和成像系统对测量肺磨玻璃密度结节CT体积准确性的影响 被引量:6

Impact of radiation dose and imaging system on the volumetric measurement accuracy of ground-glass opacity nodules
下载PDF
导出
摘要 目的评价辐射剂量和成像系统对测量磨玻璃密度(GGO)结节体积准确性的影响。方法应用MDCT和宝石能谱CT(HDCT)在7种不同噪声指数(NI)下对含有14个相同大小GGO结节的胸部体模进行扫描,采用0.625mm层厚、骨算法重建获得图像,使用Lung VCAR软件测量GGO结节体积。根据CT机器测定的剂量指数(CTDI)计算每次扫描的有效辐射剂量(ED)。应用方差分析和配对t检验对所测结节体积进行分析。结果 14个结节标准体积均为981.25mm3,7种不同NI扫描时MDCT和HDCT检测的结节体积间差异有统计学意义(P均<0.05)。随着NI增加,MDCT和HDCT测得结节体积的误差率逐渐增大,而ED逐渐下降。相同NI下,宝石能谱CT的ED较MDCT降低34.48%~50.00%,差异有统计学意义。结论相同NI下,HDCT测量GGO体积的准确性优于MDCT。在NI=15时,HDCT的ED较低,测量GGO结节体积的误差较小。 Objective To evaluate the impact of the radiation dose and imaging system on the volumetric measurement accuracy of ground-glass opacity(GGO) nodules detection.Methods MDCT and HDCT scan were performed on chest phantom containing 14 identical artificial GGO nodules with 7 different noise indexes(NI=10,15,20,25,30,35,40).All data were reconstructed with slice thickness of 0.625 mm and bone algorithm for reconstruction.Nodule volumes were measured using Lung VCAR software.Effective dose(ED) was calculated on the basis of dose length product and volumetric CT dose index.Analysis of variance and paired t test were used to analyze the measured nodule volumes.Results The standard volume of all artificial GGO nodules was 981.25 mm3.There were statistically significant differences between the groups of 7 different NI for the error of volumetric measurement of GGO nodules by MDCT and HDCT(all P0.05).With the increase of NI,nodule volume error enlarged gradually measured with MDCT and HDCT,while ED gradually decreased.ED for the chest scan on HDCT was significantly lower than that on MDCT with 34.48%—50.00% in the same NI,and statistical difference was found between them.Conclusion In the same NI,the volume measurement accuracy of HDCT is superior to volume CT.When NI of 15,HDCT scan can not only ensure low ED,but also reduce the error of measurement on nodule volume.
出处 《中国医学影像技术》 CSCD 北大核心 2011年第12期2385-2388,共4页 Chinese Journal of Medical Imaging Technology
关键词 体层摄影术 X线计算机 辐射剂量 磨玻璃密度结节 噪声指数 Tomography X-ray computed Radiation dosage Ground-glass opacity nodules Noise index
  • 相关文献

参考文献16

  • 1Goodman LR, Gulsun M, Washington L, et al. Inherent variabil- ity of CT lung nodule measurements in vivo using semiautomated volumetric measurements. AIR Am J Roentgenol, 2006,186(4) : 989-994.
  • 2Bolte H, Riedel C, Jahnke T, et al. Reproducibility of computer- aided volumetry of artificial small pulmonary nodules in ex vivo porcine lungs. Invest Radiol, 2006,41 (1) : 28-35.
  • 3Cody DD, Kim H J, Cagnon CH, et al. Normalized CT dose index of the CT scanners used in the National Lung Screening Trial. AJR Am J Roentgenol, 2010,194(6) :1539-1546.
  • 4Harrison J, Day P. Radiation doses and risks from internal emit- ters. J Radiol Prot, 2008,28(2) : 137-159.
  • 5Min JH, Lee HY, Lee KS, et al. Stepwise evolution from a focal pure pulmonary ground-glass opacity nodule into an invasive lung adenocarcinoma: an observation for more than 10 years. Lung Cancer, 2010,69(1):123-126.
  • 6Ravenel JG, Leue WM, Nietert PJ, et al. Pulmonary nodule vol- ume: effects of reconstruction parameters on automated measure- ments--a phantom study. Radiology, 2008,247(2):400-408.
  • 7Gavrielides MA, Zeng R, Kinnard LM, et al. Information-theo- retic approach for analyzing bias and variance in lung nodule size estimation with CT: a phantom study. IEEE Trans Med Ima- ging, 2010,29(10) :1795-1807.
  • 8Sone S, Tsushima K, Yoshida K, et al. Pulmonary nodules: Pre- liminary experience with semiautomated volumetric evaluation by CT stratum. Acad Radiol, 2010,17(7):900-911.
  • 9Gavrielides MA, Kinnard LM, Myers KJ, et al. Noncaleified lung nodules: volumetric assessment with thoracic CT. Radiolo- gy, 2009,251(1) :26-37.
  • 10Gavrielides MA, Kinnard LM, Myers KJ, et al. A resource for the assessment of lung nodule size estimation methods: database of thoracic CT scans of an anthropomorphic phantom. Opt Ex- press, 2010,18(14) : 15244-15255.

二级参考文献14

  • 1魏宝杰,张镭,翟仁友,王亚杰.低剂量双螺旋CT薄层扫描筛查早期肺癌初步研究[J].中国医学影像技术,2003,19(11):1519-1522. 被引量:27
  • 2Roberts HC,Patsios D,Paul NS,et al.Lung cancer screening with low-dose computed tomography:Canadian experience.Can Assoc Radiol J,2007,58(4):225-235.
  • 3Itoh S,Ikeda M,Mori Y,et al.Lung:feasibility of a method for changing tube current during low-dose helical CT.Radiology,2002,224(3):905-912.
  • 4Rampado O,Marchisio F,Izzo A,et al.Effective dose and image quality evaluations of an automatic CT tube current modulation system with an anthropomorphic phantom.Eur J Radiol,2009,72(1):181-187.
  • 5Papadakis A,Perisinakis K,Damilakis G.Angular on-line tube current modulation in multidetector CT examinations of children and adults:the influence of different scanning parameters on dose reduction.Med Phys,2007,34(7):2864-2874.
  • 6Menke J.Comparison of different body size parameters for individual dose adaptation in body CT of adults.Radiology,2005,236(2):565-571.
  • 7Kalra MK,Maher MM,Toth TL,et al.Techniques and applications of automatic tube current modulation for CT.Radiology,2004,233(3):649-657.
  • 8Bach PB,Silvestri GA,Hanger M,et al.Screening for lung cancer:ACCP evidence-based clinical practice guidelines (2nd ed).Chest,2007,132(3 Suppl):69S-77S.
  • 9Swensen SJ,Jett JR,Hartman TE,et al.CT screening for lung cancer:five-year prospective experience.Radiology,2005,235(1):259-265.
  • 10New York Early Lung Cancer Action Project Investigators.CT Screening for lung cancer:diagnoses resulting from the New York Early Lung Cancer Action Project.Radiology,2007,243(1):239-249.

共引文献52

同被引文献54

  • 1宋伟,金征宇,严洪珍,王沄,张云庆,王林辉,朱海峰,梁继祥,齐兵.初步评估16层螺旋CT的Lung Care软件在肺结节研究中的辅助价值[J].中华放射学杂志,2005,39(1):11-16. 被引量:24
  • 2王华,陈克能.细支气管肺泡癌的研究进展[J].国际呼吸杂志,2007,27(8):591-594. 被引量:9
  • 3李铁一.中华影像医学:呼吸系统卷[M].北京:人民卫生出版社,2002.238—239.
  • 4Herzog C, Mulvihill DM, Nguyen SA, et al. Pediatric cardiovas- cular CT angiography: Radiation dose reduction using automatic anatomic tube current modulation. AJR Am J Roentgenol, 2008, 190(5) : 1232-1240.
  • 5Lee T, Tsai IC, Fu YC, et al. Using multidetector-row CT in ne- onates with complex congenital heart disease to replace diagnostic cardiac catheterization for anatomical investigation: Initial, experi- ences in technical and clinical feasibility. Pediatr Radiol, 2006,36 (12) :1273-1282.
  • 6Mayo JR, Aldrich J, Muller NL, et al. Radiation exposure at chest CT: A statement of the Fleisehner Society. Radiology, 2003,228(1) : 15-21.
  • 7Heyer CM, Mohr PS, Lemburg SP, et al. Image quality and ra- diation exposure at pulmonary CT angiography with 100- or 120- kVp protocol: prospective randomized study. Radiology, 2007, 245(2) : 577-583.
  • 8Thomas KE, Wang B. Age-specific effective doses for pediatric MSCT examinations at a large children's hospital using DLP con- version coefficients: A simple estimation method. Pediatr Radiol, 2008,38(6) : 645-656.
  • 9Sigal-Cinqualbre AB, Hennequin R, Abada HT, et al. Low-ki- lovoltage multi-detector row chest CT in adults: Feasibility and effect on image quality and iodine dose. Radiology, 2004, 231 (1) :169-174.
  • 10Mccollough CH, Schueler BA. Calculation of effective dose. Med Phys, 2000,27(5) :828-837.

引证文献6

二级引证文献126

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部