高压灭菌对不同镍钛锉抗扭力性能和切割效力的影响 被引量:3

Effect of sterilization on the torque resistance and cutting efficiency of two different types of nickel-titanium rotary instruments
原文传递
导出
摘要 目的评价反复高压灭菌对两种镍钛锉抗扭力性能和切割效率的影响。方法新的ProTaperF1锉和Mtwo20#锉各60支,分为3组:对照组(未高压灭菌)、A组(连续高压灭菌7次)和B组(连续高压灭菌14次)。分别随机取3组中ProTaperF1锉和Mtwo20#锉各10支进行抗扭力性能试验,测定其在同转速和同扭距下的折断时间;分别取3组中ProTaperF1锉和Mtwo20#锉另外各10支进行切割效率测定。结果 A、B组与对照组的ProTaperF1锉和Mtwo20#锉在扭力作用的折断时间差异均无统计学意义(P>0.05);与对照组相比,A组的ProTaperF1锉和Mtwo20#锉切割力分别下降20.34%和15.34%,B组的ProTaper F1锉和Mtwo20#锉切割力分别下降46.39%和35.89%。结论高压灭菌并未增加镍钛锉折断的可能性,但可以降低镍钛锉的切割效率。 Objective To evaluate the torque resistance and cutting efficiency of two different types of nickel-titanium rotary instruments under repeated sterilization cycles. Methods 60 new ProTaper F1 and 60 new Mtwo 20# were divided into 3 groups. Each group has 20 ProTaper F1 and 20 Mtwo 20# respectively. Samples in group A were exposed to 7 consecutive sterilization cycles in a steam autoclave. Samples in group B were exposed to 14 consecutive sterilization cycles; the control group samples were not sterilized. The torque resistance and cutting efficiency of ProTaper F1 and Mtwo 20# of each group were tested. Results Compared with the control group, the time of fracture of instruments under repeated sterilization cycles was shortened, but there is no significant difference (P〉0.05, ANONA test); the cutting efficiency of ProTaper F1 and Mtwo 20# in group A were decreased by 20.34% and 15.34% respectively, while those in group B were decreased by 46.39% and 35.89% respectively. Each of these differences was statistically significant (P〈0.05, ANONA test). Conclusions Repeated sterilization in autoclave does not increase the likelihood of instrument fracture, but reduces the cutting efficiency of nickel-titanium rotary instruments.
出处 《中华口腔医学研究杂志(电子版)》 CAS 2011年第6期18-20,共3页 Chinese Journal of Stomatological Research(Electronic Edition)
关键词 灭菌 镍钛锉 扭力 切割效率 Sterilization Nickel-titanium instruments Torque Cutting efficiency
  • 相关文献

参考文献8

  • 1Roland DD,Andelin WE,Browning DF,et al.The effect of preflaring on the rates of separation for 0.04 taper nickel titanium rotary instruments.J Endod,2002,28(7):543-545.
  • 2Rapisarda E,Bonaccorso A,Tripi TR,el al.Effect of sterilization on the cutting efficiency of rotary nickel-titanium endodontic files.Oral Surg Oral Med Oral Pathol Oral Radiol Endod,1999,88(3):343-347.
  • 3孙炜,侯本祥.3种镍钛机用根管预备器械抗扭力性能的比较[J].华西口腔医学杂志,2010,28(5):471-474. 被引量:8
  • 4Berutti E,Negro AR,Lendini M,et al.Influence of manual preflaring and torque on the failure rate of ProTaper rotary instruments.J Endod,2004,30(4):228-230.
  • 5Morrison SW,Newton CW,Brown CE Jr.The eriects of stem sterilization and usage on cutting efficiency of endodontic instruments.J Endod,1989,15(9):427-431.
  • 6Viana AC,Conzaleg BM,Buono VT,el al.Iniluence of sterilization on mechanical properties and fatigue resistance of nickel-titanium rotary endodontic instruments.Int Endod J,2006,39(9):709-715.
  • 7Hilfer PB,Bergeron BE,Mayerchak MJ,el a1.Multiple autoclave cycle effects on cyclic fatigue of nickel-titanium rotary files produced by new manufacturing methods.J Endod,2011,37(1):72-74.
  • 8Ebihara A,Yabata Y,Miyara K,et a1.Heat treatment of nickel-titanium rotary endodontie instruments:effects on bending properties and shaping abilities.Int Endod J,2011,44(9):843-849.

二级参考文献18

  • 1侯本祥,徐爱凤,张琛,凌建军.超声工作尖取根管内金属折断器械的效果评价[J].北京口腔医学,2006,14(2):120-122. 被引量:31
  • 2袁理,岳林,王嘉德.Hero642镍钛锉断裂损伤的形态研究[J].实用口腔医学杂志,2006,22(5):604-607. 被引量:20
  • 3崔约贤,王长利.金属端口分析[M].哈尔滨:哈尔滨工业大学出版社,1998.
  • 4Civjan S,Hnget EF,DeSimon LB.Potential applications of certain nickel-titanium(nitinol) alloys[J].J Dent Bes,1975,54(1):89-96.
  • 5Walia HM,Brantley WA,C,erstein H.An initial investigation of the bending and torsional properties of Nitinol root canal files[J].J Endod,1988,14(7):346-351.
  • 6Yared G.In vitro study of the torsional properties of new and used ProFile nickel titanium rotary files[J].J Endod,2004,30 (6):410-412.
  • 7Pruett JP,Clement DJ,Cames DL Jr.Cyclic fatigue testing of nickel-titanium endodontic instmments[J].J Endod,1997,23 (2):77-85.
  • 8Sattapan B,Nervo GJ,Palamara JE,et al.Defects in rotary nickel-titanium files after clinical use[J].J Endod,2000,26(3):161-165.
  • 9Ullmann CJ,Peters OA.Effect of cyclic fatigue on static fracture loads in ProTaper nickel-titanium rotary instruments[J].J Endod,2005,31 (3):183-186.
  • 10Kitchens GG Jr,Liewehr FR,Moon PC.The effect of opera-.

共引文献7

同被引文献26

引证文献3

二级引证文献15

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部