期刊文献+

近红外光谱分析中建模光谱宽度的选择 被引量:6

Selection of spectral width for prediction modeling in near-infrared spectroscopy analysis
下载PDF
导出
摘要 为选择生化样品近红外光谱分析中最优的建模光谱宽度,结合样品特征波长与光谱学原理开展研究.收集348份血清样品,以血清中胆固醇、甘油三酯为例进行分析.根据光谱学原理及2 mm厚度血清的光谱吸光度,选择一级倍频区域作为分析波段,考察该区域内不同宽度光谱范围的模型效果.优选出胆固醇与甘油三酯的最优光谱宽度分别为70 nm和100 nm,并建立相应分析模型,分析的预测标准差和相对预测误差分别是0.17 mmol/L,3.0%和0.14 mmol/L,10.3%,效果不弱于他人建立模型的最好结果,使用的光谱范围更窄. In order to select optimal spectral width for biomedical components analysis by near-infrared(NIR) spectroscopy,starting with the characteristic wavelengths of the samples,a modeling procedure based on spectroscopy principles is presented.348 human serum samples are collected.Cholesterol and triglyceride in human sera are analyzed as an example.According to spectroscopy principles and absorbance of 2 mm-thick sera,first overtone region is selected for analysis.Models with different spectral width are compared.For cholesterol and triglyceride models,optimal spectral width is 70nm and 100nm respectively.Root mean square error of prediction(RMSEP) and mean percent error of prediction(MPEP) are 0.17 mmol/L,3.0% and 0.14mmol/L,10.3%,respectively.The performance is comparable with the best works of other groups while the spectral width is narrower.
出处 《红外与毫米波学报》 SCIE EI CAS CSCD 北大核心 2011年第6期522-525,共4页 Journal of Infrared and Millimeter Waves
基金 国家自然科学基金(60878052,60938002) 应用光学国家重点实验室开放基金(O9Q13FQ090) 广东省自然科学基金(8151063201000017)
关键词 近红外光谱 血清成分测量 偏最小二乘 建模谱宽选择 near-infrared spectroscopy examination of serum component partial least squares(PLS) selection of spectral width
  • 相关文献

参考文献14

  • 1Arnold M A, Small G W. Noninvasive glucose sensing[ J]. Analytical Chemistry,2005,77 ( 17 ) :5429 - 5439.
  • 2陈星旦.近红外光谱无创生化检验的可能性[J].光学精密工程,2008,16(5):759-763. 被引量:51
  • 3丁海泉,卢启鹏,朴仁官,陈星旦.土壤有机质近红外光谱分析组合波长的优选[J].光学精密工程,2007,15(12):1946-1951. 被引量:35
  • 4高洪智,卢启鹏,丁海泉,彭忠琦.基于连续投影算法的土壤总氮近红外特征波长的选取[J].光谱学与光谱分析,2009,29(11):2951-2954. 被引量:59
  • 5Heise H M, Marbach R. Multivariate determination of blood substrates in human plasma by FT-NIR spectroscopy [ J ]. Proc. SPIE, 1992,1575:507 - 508.
  • 6Chen J, Arnold M A, Small G W. Comparison of combination and first overtone spectral regions for near-infrared calibration models for glucose and other biomolecules in aqueous solutions [ J ]. Analytical Chemistry, 2004,76 ( 18 ) : 5405 - 5413.
  • 7Hazen K H, Arnold M A, Small G W. Measurement of glucose and other analytes in undiluted human serum with near-infrared transmission spectroscopy [ J ]. Analytica Chimica Acta, 1998,371 (2 - 3 ) :255 - 267.
  • 8陈华才,杨仲国,李惠英,陈星旦.人血清中胆固醇近红外光谱快速检测初步研究[J].激光生物学报,2004,13(6):429-432. 被引量:2
  • 9Peuchant E, Salles C, Jensen R. Determination of serum cholesterol by near-infrared reflectance spectrometry [ J ]. Analytical Chemistry, 1987,59( 14 ) : 1816 - 1819.
  • 10Haaland D M, Robinson M R, Koepp G W, et al. Ragenfless near-infrared determination of glucose in whole blood using muhivariate calibration[ J]. Applied Spectroscopy, 1992,46 (10) : 1575 - 1578.

二级参考文献37