电感耦合等离子体原子发射光谱法测定硅锰合金中的铝 被引量:1

DETERMINATION OF ALUMINUM IN SILICON-MANGANESE ALLOY WITH INDUCTA-NCE-COUPLING PLASMA ATOMIC EMMISION SPECTRUM
下载PDF
导出
摘要 提出了电感耦合等离子体原子发射光谱法测定硅锰合金中铝的测定方法。试样于铂皿中,加入硝酸氢氟酸溶解试样,高氯酸驱赶氟离子,用盐酸溶解盐类,选择波长309.3 nm为主分析线,方法的检出限为0.02 mg/L,分析了几批试样,测定值与国家标准方法测定值相对误差<0.22%,回收率在98.67%~100.14%。 It is stated the method to determine aluminum in silicon - manganese alloy with induction - coupling plasma atomic emission spectrum: put sample in a platinum pot, add into nitric acid plus hydrofluoric acid to dissolve the sample, perchloric acid to drive away fluorine ion, hydrochloric acid to dissolve salt, select 309. 3 nm wavelength as main analysis line. The check - out limit is 0.02 mg/L. With several groups of sample analyzed, the relative error between determined values and ones got with national standard method is lower than 0.22% ; the recovery ratio is between 98.67 and 100.14 percent.
出处 《河北冶金》 2011年第12期24-24,78,共2页 Hebei Metallurgy
关键词 电感耦合等离子体 硅锰合金 测定 induction - coupling plasma silicon - manganese alloy determination aluminum
  • 相关文献

参考文献2

  • 1辛仁轩.等离子体发射光谱分析[M].北京:化学工业出版社,2006:73-104.
  • 2孙维贞.常见元素的化学分析方法[M].北京:化学工业出版社,2008:21-26.