期刊文献+

血清总蛋白测定的基质效应评价 被引量:2

Evaluation of matrix effect for serum total protein
下载PDF
导出
摘要 目的评价14种室间质量评价(EQA)材料和3种市售材料在4个检测系统上测定血清总蛋白的基质效应。方法参照美国临床实验室标准化协会(CLSI)EP14-A2指南,以美国临床化学协会(AACC)推荐血清总蛋白双缩脲为参考方法,以4个检测系统为待评方法,评价制备物的基质效应。结果 9种制备物对各系统均无基质效应,1种制备物对所有系统均存在基质效应,其余样本对不同检测系统存在不同程度的基质效应。结论制备物样本的基质效应可影响血清总蛋白测定的准确性和可比性,在临床检测中应重视其对检验结果量值溯源的影响。 Objective To evaluate the matrix effects of measurement for total protein in serum using 14 materials of external quality assessment(EQA) and 3 commercial reference materials in 4 measurement systems. Methods The matrix effects of 17 processed samples were evaluated by EP14 - A2. The biuret method for serum total protein, recommended by the American Association for Clinical Chemistry(AACC), was reproduced as reference method. 4 measurement systems were evaluated as compared method. Results Evaluation of matrix effect: nine processed samples showed no matrix effect at all measurement systems. One processed sample Showed matrix effect at all systems. Others showed different matrix effects at dif- ferent systems. Conclusion The accuracy and comparability for serum total protein measurement is affected by the matrix effects of processed samples.
出处 《中国医学创新》 CAS 2011年第36期12-15,共4页 Medical Innovation of China
基金 国家科技支撑计划(编号:2007BAI05B09)
关键词 血清总蛋白 双缩脲反应 参考方法 基质效应 Serum total protein Biuret reaction Reference method Matrix effect
  • 相关文献

参考文献8

二级参考文献50

共引文献94

同被引文献12

引证文献2

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部