EPID在盆腔肿瘤放疗中的摆位误差分析 被引量:7

EPID在盆腔肿瘤放疗中的摆位误差分析
下载PDF
导出
摘要 目的探讨电子射野影像系统在盆腔肿瘤放疗中的摆位误差,以供临床参考。方法对17例盆腔肿瘤接受放疗的患者给予摆位误差的量化测定。所有病例给予CT的定位和放疗。将治疗过程中的电子射野影像系统所拍摄的正位和侧位片,与数字化重建图像的验证片进行对比分析,计算误差。结果 17例患者在每日摆位误差中,主要为前后的误差,在2mm~22mm之间,平均为(13.21±7.17)mm;其次为头脚,在1mm~23mm之间,平均为(8.01±5.32)mm;左右误差发生最小,在0.2mm~8.7mm之间,平均(5.27±3.99)mm。结论使用电子射野影像系统能够有效地对盆腔肿瘤放疗期间的摆位误差进行测定,为放疗提供初步的参考依据。 Objective Of electronic portal imaging system in pelvic radiotherapy in cancer of the pendulum position error for clinical reference. Methods Our hospital 17 cases of pelvic tumors were treated with radiotherapy quantify positioning errors. Given to the positioning of all cases CT and radiotherapy. The treatment of electronic portal imaging system to capture the anteroposterior and lateral films, and digital image reconstruction verification films were analyzed to calculate the error. Results 17 patients in the daily positioning errors, mainly before and after the error, in the 2mm - 22mm, the average of (13.21 ± 7.17) mm; followed by the head and feet, in 1mm- 23mm with an average of (8.01 ±5.32) mm; about the smallest error occurred in the 0.2mm - 8.7mm, the average (5.27 ± 3.99) mm. Conclusion The use of electronic portal imaging system can effectively during radiotherapy of pelvic tumors were measured positioning errors for the radiation to provide a preliminary reference.
出处 《当代医学》 2012年第2期10-11,共2页 Contemporary Medicine
关键词 盆腔肿瘤 放射治疗 摆位误差 电子射野影像系统 Pelvic tumor Radiation therapy Positioning error Electronic portal imaging system
  • 相关文献

参考文献6

二级参考文献26

  • 1鄂丽萍,李兵,熊晓斌.三维适形放射治疗患者的固定技术与摆位[J].医学研究生学报,2004,17(11):1042-1043. 被引量:15
  • 2王艳阳,傅小龙.CT在影像引导下放疗中应用的历史与现状[J].中国癌症杂志,2006,16(6):448-453. 被引量:22
  • 3邬筱莹,章真,王文超,任珺,郭晓梅,陆惠忠.电子射野影像装置对盆腔肿瘤放疗摆位误差的测定[J].中华放射肿瘤学杂志,2007,16(1):52-54. 被引量:26
  • 4Prescribing recording and reporting photon bean1 therapy(Supplement to ICRU report 50)[R]. ICRU. Report no.62,1999.
  • 5Van Herk M, Remeijer P. Lebesque jr. Inclusion of geometric uncertainties in treatment plan evaluation [J]. Int J Radiat Oncol Biol Phys, 2002,52(5): 1407-1422.
  • 6Mckenzie A, Van Herk M, Mijnheer B. Margins for geometric uncertainty around organs at risk in radiotherapy [J]. Radiotherapy and Oncology, 2002,62(3):299-307.
  • 7Hurkmans CW, Remeijer P'Lebesque JV, et al. Set-up verification using portal image review of current clinical practice [J]. Radiotherapy and Oneology, 2001,58(2):105-120.
  • 8Stroom JC, de Boer HC, Huizenga H, et al. Inclusion of geometrical uncertainties in radiotherapy treatment plan ning by means of coverage probability[J]. Int J Radiat Oncol Biol Phys, 1999,43(4):905-919.
  • 9Dawson LA, Jaffray DA. Advances in image-guided radiation therapy. J Clin Oncol, 2007,25(8) :938-946.
  • 10Herman MG. Clinical use of electronic portal imaging. Semin Radiat Oncol, 2005, 15(3) : 157-167.

共引文献33

同被引文献52

引证文献7

二级引证文献27

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部