期刊文献+

基于红外反射光容积脉搏波的血管阻力研究 被引量:6

A Research of Vascular Resistance Based on Reflected Infrared Photoplethysmography
原文传递
导出
摘要 近红外光谱技术在血氧饱和度和血糖浓度的无损伤在体测量中得到了应用。本文利用基于反射式原理的红外光容积脉搏波对指尖和桡动脉进行了血管阻力等血流参数检测的研究。通过将提取到的光电容积脉搏波中峰峰值和特征K值参数与超声多普勒所测收缩期峰值血流速度和血管阻力参数进行相关性分析,发现两者存在很强的相关性,研究结果将扩展光电容积脉搏波在血流检测中的应用。 Near-infrared spectrum technology can fulfill the non-invasive in-vivo measure- ments of oxygen saturation of blood and blood glucose concentration. This program is the research of vascular resistance based on photoplethysmography. Comparing the parameters of photoplethysmography with parameters of ultrasonic, we find there do exist an strong correlation. Compared with ultrasonic measurement, photoplethysmography do not need expensive devices, and it has many advantages such as convenient, stable performance, non-invasive.
出处 《光散射学报》 北大核心 2011年第4期406-410,共5页 The Journal of Light Scattering
关键词 血管阻力 容积脉搏波描记法 超声多普勒 相关性分析 vascular resistance photoplethysmography ultrasonic correlation
  • 相关文献

参考文献6

  • 1LIU Yuliang. An In-vivo acquisition device for near infrared blood spectra[J]. Proceedings of the 4th IEEE-EMBS International Summer School and Symposium on Medical Devices and Biosen- sors, 2007: 73--76.
  • 2Gangmin Ning. New approaches to physiological study by telemetry technology[J]. Proceedings of 2005 IEEE Engineering in Medicine and Biology 27th Annual Conference, 2005, 6654--6657.
  • 3陈受田.腹部肿瘤超声诊断计术[M].重庆:重庆出版社,2004,17-48.
  • 4Millasseau S C, Kelly R P, Ritter J M, etal. De- termination of age-related increases in large artery stiffness by digital pulse contour analysis [J]. Clinical Science, 2002, 103: 2002, 371--377.
  • 5高博,魏蔚,龚敏,王丽.脉搏血氧饱和度检测仪的研制[J].微计算机信息,2009,25(23):12-14. 被引量:6
  • 6刘晓石.概率论与数理统计[M].北京:科学出版社,2005.85-89.

二级参考文献8

  • 1刘进,魏蔚.经气管肺动脉血氧饱和度和(或)压力监测导管[P],中国:ZL031352227,2007,7,18.
  • 2Yasuhiro Saeki, Komin Takamura, Kazushige, et al. The measurement technique of human's bio-signals [C], Proceeding of the 28th IEEE EMBS Annual International Conference New York City, USA,Aug 30-Sept3,2006,1351-1354.
  • 3Wei Wei ,Zhaoqiong Zhu, Lunxu Liu, et al. A pilot Study of Continuous Transtracheal Mixed Venous Oxygen Saturation Monitoring[J], Anesth Analog 2005,101:440-44.
  • 4Ugnell H and Oberg PA, the tine-variable photoplethysmographic signal: dependence of the heart synchronous signal on wavelength and sample volume, Med.Eng.Phys.,17,571-578,1995.
  • 5Altera Corp., Cyclone Device Handbook[K], May 2003.
  • 6K. Shafqat, D.P.Jones, R.M.Langford, et al. Filtering techniques for the removal of ventilator artefact in oesophageal pulse oximetry [J], Med Bio Eng Comput,21 July 2006,44:729-737.
  • 7杨夏,王平,张军.基于FPGA的简化UART电路设计[J].微计算机信息,2008,24(23):226-227. 被引量:12
  • 8黄岚,田丰华,丁海曙,王广志,腾铁超.用近红外光谱对组织氧测量方法的研究[J].红外与毫米波学报,2003,22(5):379-383. 被引量:35

共引文献12

同被引文献48

  • 1罗志昌,张松,杨文鸣,杨子彬.脉搏波波形特征信息的研究[J].北京工业大学学报,1996,22(1):71-79. 被引量:119
  • 2乜国荃,方祖祥.人体脉搏的测量与分析[J].上海生物医学工程,2006,27(2):74-76. 被引量:17
  • 3朱翔宇,韩红远.LCD液晶数字显示红外脉搏表[J].太原科技,2006(12):79-80. 被引量:2
  • 4Vasapollo B,Novelli GP,Gagliardi G. Medical treatment of early-onset mild gestational hypertension reduces total peripheral vascular resistance and influences maternal and fetal complications[J].{H}Ultrasound in Obstetrics & Gynecology,2012,(03):325-331.
  • 5Andrew R,Phillip AS,Devin MC. Utility of the Photoplethys-mogram in Circulatory Monitoring[J].{H}ANESTHESIOLOGY,2008.950-958.
  • 6Meir N,Amikam P,Zehava G. Automatic noninvasive mea-surement of systolic blood pressure using photoplethysmography[J].{H}BioMedical Engineering OnLine,2009.28-36.
  • 7Glauco AW,Eliezer S,Anderson RG. Pulse oximetry wave variation as a noninvasive tool to assess volume status in cardiac surgery[J].CLINICS,2009,(04):337-343.
  • 8Phillips JP,Hickey M,Kyriacou PA. Evaluation of electrical and optical plethysmography sensors for noninvasive monitoring of he-moglobin concentration[J].Sensors(Basel),2012,(02):1816-1826.
  • 9Keisuke M,Kazuo AO,Naoto K. Fundamental Study of Reflec-tion Pulse Oximetry[J].{H}OPTICAL REVIEW,2003,(05):482-487.
  • 10Mou L,Gong Q,Wei W. The analysis of transesophageal oxygen saturation photoplethysmography from different signal sources[J].{H}Journal of Clinical Monitoring and Computing,2013,(03):365-370.

引证文献6