期刊文献+

新型环金属铱配合物-硫酸铈体系化学发光法测定亚硫酸盐 被引量:1

Chemiluminescence determination of sulfite using the novel cyclometalated iridium complex-sulfite-cerium sulphate system
下载PDF
导出
摘要 酸性环境中利用合成的新型环金属铱配合物[(dpci)2Ir(pcp)](dpci=3,4-二苯基邻二氮杂萘,pcp=2-(3-吡唑并[4,5]樟脑)基吡啶)-亚硫酸盐-硫酸铈反应建立了一个快速、灵敏的测定亚硫酸盐的新体系。实验结果表明,化学发光强度与亚硫酸盐浓度在3.0×10-7~1.0×10-4 mol/L范围内呈线性关系,检测限为9.8×10-8 mol/L。9次重复测定1.0×10-5 mol/L亚硫酸盐相对标准偏差为3.7%。该法用于测定白糖中的亚硫酸盐,结果令人满意。 A rapid and sensitive chemiluminescence (CL) flow system for determination sulfite is described based on the reaction of cyclometalated iridium complex [ ( dpci )2Ir- ( pcp ) ] ( dpci = 3,4-dipheny, pcp = 2- ( 3-pyrazol [ 4,5 ] cam- phor) pyridine), sulfite and cerium sulphate. The optimum conditions for the CL emission have been established. The CL emission intensity is linear with sulfite concentration in the range from 3.0 × 10-7 to 1.0 × 10-4 mol/L. The detection limit is 9. 8 ×10 -8 mol/L and the relative standard deviation is 3.7% for 1.0 × 10 -5 mol/L sulfite in nine repeated measurements. The method has successfully been used for the determination of sulfite in white sugar samples.
出处 《化学试剂》 CAS CSCD 北大核心 2012年第1期48-50,80,共4页 Chemical Reagents
基金 国家自然科学基金资助项目(50903001,5097-5002) 安徽省教育厅自然科学重点基金资助项目(KJ2011-A054) 安徽工业大学青年基金资助项目(QZ200907和201005)
关键词 环金属铱配合物 化学发光法 亚硫酸盐 测定 cyclometalated iridium complex chemilumines-cence sulfite determination
  • 相关文献

参考文献14

  • 1NAVARRRO M V,PAYAN M R,LOPEZ M A B,et al. Rapid flow injection method for the determination of sul- rite in wine using the permanganate-luminol luminescence system [ J ]. Talanta, 2010,82 ( 5 ) :2 003-2 006.
  • 2BONIFACIO R L,COICHEV N. Chemiluminescent deter- mination of sulfite traces based on the induced oxidation of Ni ( I1 )/tetraglycine complex by oxygen in the pres- ence of luminol: mechanistic considerations [ J ]. Anal. Chim. Acta, 2004,$17 ( 1/2 ) : 125-130.
  • 3LI Bao-xin, ZHANG Zhu-jun, WU Man-li. Flow-injection chemiluminescence determination of sulfite using on-line electrogenarated silver ( II ) as the oxidants [ J ]. Anal. Chim. Acta,2001,432( 2 ) :311-316.
  • 4QIN Wei, ZHANG Zhu-jun, ZHANG Cheng-jie. Chemilu- minescence flow system for the determination of sulfite [ J ]. Fresenius J. Anal. Chem. , 1998,361 : 824-826.
  • 5STAUFF J, JAESCHKE W. Determination of epoxides in automotive emissions by chemiluminescence [ J ]. Atmos Environ. , 1975,9 ( 11 ) : 1 038-1 039.
  • 6HUANG Yu-ming, ZHANG Chao, ZHANG Xin-rong, et al. A sensitive chemiluminescence flow system for the de- termination of sulfite [ J ]. Anal. Left. , 1999,32 ( 6 ) : 1 211-1 224.
  • 7YAMADA M, NAKADA T, SUZUKI S. The determination of sulfite in a flow-injection system with chemilumines- cence detection [ J ]. Anal. Chim. Acta, 1983,147 : 401- 404.
  • 8PAULLS D A,TOWNSHEND A. Sensitive determina- tion of sulflte using flow-injection with chemiluminescenct detection [ J ]. Analyst, 1995,120 ( 2 ) :467-469.
  • 9YERSIN H. Highly efficient OLEDs with phosphores- cent materials[ M ]. Yersin H. , Ed. ; Wiley-VCH, : Wein- heim, Germany, 2008.
  • 10AMAO Y,ISHIKAWA Y, OKURA I. Green luminescent iridium( Ⅲ ) complex immobilized in fluoropolymer film as optical oxygen-sensing material [ J ]. Anal. Chim. Ac- ta ,2001,445 ( 2 ) : 177-182.