电离室矩阵用于加速器质量保证:X射线 被引量:5

Application of Ionization Chamber Array in Quality Assurance of Linear Accelerator:X Ray
下载PDF
导出
摘要 目的:探讨利用二维电离室矩阵MatriXX进行医用直线加速器质量保证的方法及可靠性。方法:在点剂量的质量保证中,将MatriXX系统在特制水箱中测得的数据与0.6 cc电离室在标准水箱中的剂量测量值作比较,从而定出Ma-triXX的校准系数,以此来校准射线的输出量和能量。在面剂量的质量保证中,用MatriXX测量医用加速器射线野的对称性、平坦度以及辐射野的大小及中心轴偏离,将测量结果与RFA300三维水箱扫描结果及控制台显示数值进行比较分析。结果:用MatriXX加有机玻璃板可准确快捷地完成对射线输出量和能量的校准。用MatriXX所测射野的对称性和平坦度及射线野大小与RFA300水箱的测量数据一致。结论:MatriXX系统进行加速器的质量保证简便可靠。 Objective: To investigate the method and feasibility of 2D ionization chamber array MatriXX system in quality assurance of linear accelerator. Methods: The outputs and the energy levels irradiated for phonton beams measured by MatriXX in a water phantom specially designed were compared with the data acquired by 0.6 cc ionization chamber with the standard water phantom. The comparison fixed on a correction factor which was used to check the output and energy. The symmetry and flatness and field sizes of the photon beam measured by MatriXX system were compared with the data acquired by RFA300 3D scanner and the digit display of field size showed in the console. Results: The Output arid the energy level are calibrated accurately and quickly by MatriXX system. The symmetry, flatness and field sizes of the beam measured by MatriXX system are coincident with the data acquired by RFA300 3D scanner and the data showed in the console. Conclusions: The MatriXX system is very convenient and feasible in the quality assurance of linear accelerator.
机构地区 南方医院放疗科
出处 《中国医学物理学杂志》 CSCD 2012年第1期3096-3098,共3页 Chinese Journal of Medical Physics
关键词 剂量测量 电离室矩阵 加速器QA dosimetry MatriXX QA of linear accelerator
  • 相关文献

参考文献6

二级参考文献12

  • 1张书旭,徐海荣,陈光杰,杨克柽,林生趣,余辉.呼吸运动对靶区三维重建的影响[J].中国医学物理学杂志,2006,23(6):400-404. 被引量:14
  • 2胡杰,陶建民,孙光荣.医用直线加速器等中心的质量控制和质量保证[J].中国医疗器械杂志,2007,31(3):213-215. 被引量:24
  • 3瞿福山 封三友 等.直线加速器机械及光学系统的检测[J].中华放射医学与防护杂志,1995,15(3):161-161.
  • 4Andenna C, Benassi M, Caccia B, et al. Comparison of dose distributions in IMRT planning using the gamma function [ J ]. J Exp Clin Cancer Res,2006,25 (2) :229 - 234.
  • 5Amerio S, Boriano A, Bourhaleb F, et al. Dosimetric characterization of a large area pixel - segmented ionization chamber [J]. MedPhys,2004,31(2):414-420.
  • 6Herzen J, Todorovic M, Cremers F, et al. Dosimetric evaluation of a 2D pixel ionization chamber for implementation in clinical routine[J]. Phys Med Biol,2007,52(4) : 1 197 - 1 208.
  • 7Jursinic PA, Nelms BE. A 2 - D diode array and analysis software for verification of intensity modulated radiation therapy delivery [ J ]. Med Phys, 2003,30 ( 5 ) : 870 - 879.
  • 8Wiezorek T, Banz N, Schwedas M. Dosimetric quality assurance for intensity - modulated radiotherapy feasibility study for a filmless approach [ J ]. Strahlenther Onkol, 2005, 181 (7) :468 -474.
  • 9Ametio S,Boriano A,Bourhaleb F,et al.Dosimetric characterization of a large area pixel-segmented ionization chamber[J].Med Phys.2004,31 (2) : 414-420.
  • 10Herzen J,Todorovic M,Cremers F,et al.Dosimetric evaluation of a 2D pixel ionization chamber for implementation in clinical routine[J].PhyS Med Biol.2007,52(4) : 1197-1208.

共引文献19

同被引文献32

引证文献5

二级引证文献9

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部