AX-4280全自动尿液分析仪性能评价 被引量:7

To evaluate the performance of AX-4280 automatic urine dry chemistry analyzer
下载PDF
导出
摘要 目的评价AX-4280全自动尿液分析仪的性能。方法采用AX-4280全自动尿液分析仪,分别检测各项参数的精密度、准确度、携带污染率等。结果高、低质控品精密度SD%值(0.001~0.943、0~0.832)、携带污染率(0.8%~10.0%)、准确度等均在仪器要求范围内。结论 AX-4280全自动尿液分析仪各项性能均符合仪器要求范围,可用于临床尿液干化学分析检测。 Objective To evaluate the performance of the AX-4280,which is the fully automated urine dry chemistry analyzer. Methods The AX-4280 automatic urine dry chemistry analyzer was used to detect the precision,accuracy and carryover rates of all parameters. Results The SD values of the hight-control and lowr-control precisions(0. 001-0. 943、0-0. 832) ,the contamination rate((0.8%--10.0%)and accuracy were all within the requirements of the instrument. Conclusion The performance could meet the requirements of the AX-4280 analyzer,which can be used for dry chemical analysis of clinical urine testing.
出处 《重庆医学》 CAS CSCD 北大核心 2012年第3期262-263,共2页 Chongqing medicine
关键词 尿液分析仪 性能评价 精密度 准确度 携带污染率 urine analyzer performance evaluation precision accuracy contamination rate