IDEAL序列在脊柱脂肪抑制中的应用 被引量:9

Application of Spine Fat Suppression Using IDEAL
下载PDF
导出
摘要 目的比较IDEALT2WI、FSET2WI序列在脊柱脂肪抑制扫描中的应用价值。资料与方法 35例脊柱MRI检查患者(颈椎10例、胸椎8例、腰椎17例)同时采用IDEALT2WI、FSET2WI两种脂肪抑制序列扫描,对图像脂肪抑制质量进行主观评价分级评估并测量信噪比。结果 FSET2WI脂肪抑制质量主观评价分级为颈椎(1.30±0.48)级、胸椎(1.80±0.71)级、腰椎(2.30±0.69)级,而IDEALT2WI序列颈、胸椎均为(3.00±0.00)级、腰椎为(2.80±0.56)级;FSET2WI脂肪抑制序列信噪比平均值分别为颈(8.73±4.66)、胸(11.33±9.27)、腰(6.81±10.15),IDEALT2WI序列信噪比平均值分别为颈(18.90±7.71)、胸(26.02±11.61)、腰(19.57±9.12),两序列信噪比差异有统计学意义(t=3.73、5.72、11.23,P<0.05)。结论脊柱MRI脂肪抑制扫描IDEALT2WI序列优于FSET2WI序列,IDEALT2WI序列能提供更均匀稳定的脂肪抑制效果,脊柱图像信噪比高,可清晰显示脊柱病变。 Purpose To assess the clinical value of iterative decomposition of water and fat with echo asymmetry and least-squares estimation (IDEAL) in spine imaging by comparing the image quality of T2WI with frequency-selective fat saturation. Materials and Methods Thirty-five patients (cervical 10, thoracic 8, lumber 17) underwent spine MR examination using both IDEAL T2-weighted image (T2WI) and frequency-selective fat saturation fast spin echo (FSE) T2WI , and the image quality was compared. Results The image quality of frequency-selective fat saturation on FSE T2WI were 1.30±0.48 in cervical spine, 1.80±0.71 in thoracic spine, 2.30±0.69 in lumbar spine respectively, and that of IDEAL T2WI were 3.00±0.00 in cervical spine, 3.00±0.00 in thoracic spine, 2.80±0.56 in lumbar spine respectively. The signal-to-noise ratio (SNR) of frequency-selective fat saturation on FSE T2WI were 8.73±4.66 in cervical spine, 11.33±9.27 in thoracic spine, 6.81±10.15 in lumbar spine respectively, and that for IDEAL T2WI were 18.90±7.71, 26.02±11.61, 19.57±9.12 respectively. The SNR of IDEAL T2WI was significantly different from that of frequency-selective fat saturation FSE T2WI (t =3.73, 5.72, 11.23, P 〈 0.05 ) Conclusion The image quality and SNR of IDEAL T2WI are superior to that of frequency-selective fat saturation FSE T2WI.
出处 《中国医学影像学杂志》 CSCD 北大核心 2012年第1期2-4,共3页 Chinese Journal of Medical Imaging
关键词 IDEAL序列 脊柱 脂肪抑制 磁共振成像 IDEAL sequence Spine Fat suppression Magnetic resonance imaging
  • 相关文献

参考文献6

  • 1Glover GH, Schneider E. Three-point dixon technique for true water/fat decomposition with B0 inhomogeneity correction. Magn Reson Med, 1991, 18(2): 371-383.
  • 2Xiang Q, An L. Water-fat imaging with direct phase encoding. J Magn Reson Imaging, 1997,7(6): 1002-1015.
  • 3Rybicki F J, Chung T, Reid J, et al. Fast three-point Dixon MR imaging using low-resolution images for phase correction: a comparison with chemical shift selective fat suppression for pediatric musculoskeletal imaging. AJR, 2001, 177(5): 1019-1023.
  • 4Ma J, Singh SK, Kumar AJ, et al. Method for efficient fast spin echo Dixon imaging. Magn Resort Med, 2002, 48(6): 1021-1027.
  • 5Reeder SB, Wen Z, Yu H, et al. Multicoil Dixon chemical species separation with an iterative least-squares estimation method. Magn Reson Med, 2004, 51(1): 35-45.
  • 6Reeder SB, Yu H, Johnson JW, et al. T1- and T2-weighted fast spin-echo imaging of the brachial plexus and cervical spine with IDEAL water-fat separation. J MRI, 2006, 24(4): 825-832.

同被引文献76

引证文献9

二级引证文献59

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部