摘要 工频X线机可分为常规X线机和程控X线机,常规X线机价格低廉,且能满足一般的医学临床需求,因而广泛应用于基层医院特别是乡镇一级的医院中。基层医院的大部分X线机使用年限较长,维修和保养的条件较差,发生故障后检修起来较为困难。在日常工作中对FSK132—1型X线机特殊故障和FSB202—1型X线机特殊故障的出现,经细心查找发现了问题所在。检修X线机要对故障进行综合分析,制定完整的检修计划,按计划逐步进行检查,切忌无计划的盲动修理。 X-ray machines can be divided into two types: conventional X-ray machine and the programmable X machine. The conventional X-ray machine of low price can meet the clinical demand, thus widely exists in the basic level hospitals especially the township level hospitals.In the basic level hospitals, most of the X-ray machine used for a long time, so its repair and maintenance are of poor condition. Fault maintenance is thus more difficult. The author described here 2 special breakdowns of X-ray machine for reference.
作者 寻长洲
出处 《中国医学装备》 2012年第1期81-82,共2页 China Medical Equipment
关键词 X线机 维修 故障 X-ray machine Maintenance Breakdown