期刊文献+

Excel表格在临床酶学参考实验室测量不确定度评定中的初步应用 被引量:1

下载PDF
导出
摘要 目的探讨Excel表格在临床酶学参考实验室测量不确定度评定中的应用。方法以γ-谷氨酰基转移酶(GGT)为例,用"自下而上"方法分析GGT测量不确定度的来源,并对各不确定度分量进行评定。用Microsoft Excel 2007表格记录并计算。结果 GGT的扩展不确定度为4.4 U/L。GGT的测量结果报告为:(207.4±4.4)U/L,k=2。结论 Excel表格使用方便、快速,可用于临床酶学参考实验室测量不确定度的评定。
出处 《临床检验杂志》 CAS CSCD 北大核心 2011年第9期653-654,659,共3页 Chinese Journal of Clinical Laboratory Science
  • 相关文献

参考文献6

  • 1Suzanne Castmp integrated sciences group. A comprehensive comparisonof uncertainty analysis tools [ C ]. Anaheim : Measurement Science Conference, 2004 : 1-12.
  • 2Kragten J. Calculating standard deviations and confidence intervals with a universally applicable spreadsheet technique [ J ]. Analyst, 1994,119:2161-2166.
  • 3Schumann G, Bonora R, Ceriotti F, et al. IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37 degrees C. International Federation of Clinical Chemis- try and Laboratory Medicine. Part 6. Reference procedure for the measurement of catalytic concentration of y-glutamyltransferase [ J ]. Clin Chem Lab Med,2002,40(7) :734-738.
  • 4BIPM, IEC, IFCC, ISO, IUPAC, IUPAP, OIML. Guide to the expres- sion of uncertainty in measurement (GUM) [ M ]. 1st edition. ISO, 1995.
  • 5CLSI CS1-P. Expression of measurement uncertainty in laboratory medicine. Proposed guideline [ S ]. CLSI,2010.
  • 6陈宝荣,孙慧颖,邵燕,胡滨,李月玲.基于“实验设计”的临床酶学参考方法测量不确定度评定模型的建立[J].临床检验杂志,2011,29(5):327-330. 被引量:9

二级参考文献2

  • 1BIPM, IEC, IFCC, ISO, IUPAC, IUPAP, OIML. Guide to the ex- pression of uncertainty in measurement ( GUM ) [ M ]. 1st edition. ISO, 1995.
  • 2EURACHEM/CITAC guide CG4. Quantifying uncertainty in analytical measurement (QUAM) [ M ]. 2 nd edition. EURACHEM/CITAC ,2000.

共引文献8

同被引文献12

  • 1ISO 18153. In vitro diagnostic medical devices-Measurement of quantities in biological samples-Metrological traceability of values for catalytic concentration of enzymes assigned to calibrators and control materials [S]. /SO,2003.
  • 2Infusino I, Schumann G, Ceriotti F,et al. Standardization in clinical enzymology: a challenge for the theory of metrologieal traceability [J]. Clin Chem Lab Med,2010,48(3) :301-307.
  • 3ISO/IEC Guide 98-3. Uncertainty of measurement-Part3: Guide to the expRession of uncertainty in measurement ( GUM: 1995 ) [ S ]. ISO/IEC ,2008.
  • 4JCGM 101. Evaluation of measurement data Supplement 1 to the "Guide to the expRession of uncertainty in measurement" -- Propa-gation of distributions using a Monte Carlo method [ M ]. JCGM ,2008.
  • 5Schumann G, Bonora R, Ceriotti F, et al. IFCC Primary Reference ProceduRes for the Measurement of Catalytic Activity Concentrations of Enzymes at 37 C. Part 2. Reference Procedure for the Measure- ment of Catalytic Activity Concentration of Creatine Kinase [ ATP: Creatine N-Phosphotransferase ( CK), EC 2.7. 3.2 1 [ J ]. Clin Chem Lab Med,2002,40(6) :635-642.
  • 6中国合格评定国家认可委员会.CNAS-CL33:检测和校准实验室能力认可准则在临床酶学参考测量领域的应用说明[Z].2011-09-22.
  • 7EURACHEM/CITAC Guide, Quantifying Uncertainty in Analytical Measurement[ M]. 3rd ed. London :Laboratory of the Government Chemist, 2012 : 108-111.
  • 8黄美发,景晖,匡兵,钟艳如,蒋向前.基于拟蒙特卡罗方法的测量不确定度评定[J].仪器仪表学报,2009,30(1):120-125. 被引量:36
  • 9陈宝荣,孙慧颖,邵燕,胡滨,李月玲.基于“实验设计”的临床酶学参考方法测量不确定度评定模型的建立[J].临床检验杂志,2011,29(5):327-330. 被引量:9
  • 10酶学参考实验室参考方法测量不确定度评定指南[J].临床检验杂志,2011,29(9):698-720. 被引量:6

引证文献1

二级引证文献5