期刊文献+

大气/真空环境紫外臭氧垂直探测仪光谱辐照度定标研究 被引量:17

Spectral Irradiance Calibration Research of the Solar Backscattered Ultraviolet Spectroradiometer in Air/Vacuum Conditions
原文传递
导出
摘要 为进一步提高紫外臭氧垂直探测仪(SBUS)在轨探测及数据反演精度,提出了SBUS地面定标的一系列改进方案,其中为实现地面定标与在轨工作环境一致的全波段真空辐射定标是改进的第一项。通过构建SBUS大气/真空光谱辐照度响应度比对测试装置,实测了两种环境下SBUS整机对同一光源的光谱辐照度响应。结果显示,在250~300nm波段,真空/大气相对偏差约0.8%;在300~400nm波段,真空/大气比对结果随波长变化,最大偏差略高于15%。而仪器250~400nm波段定标环境引入的单项不确定度,真空定标比以往大气定标减小了1.8%。理论分析及实验验证后发现SBUS反射元件Al+MgF2膜层在真空/大气下光谱反射率会发生变化,从而证实了SBUS在真空环境下定标的必要性。 A group of improving methods for the on-ground calibration of solar backscattered ultraviolet spectroradiometer (SBUS) has been put forward to increase its precision of on-board detection and data inversion. The full-waveband radiometric calibration in vacuum, which can attain correspondence match of condition between on-ground calibration and on-board detection, is the first item. A spectral irradiance responsibility comparing measurement setup is built and the spectral irradiance response of SBUS to a same source is measured in air and vacuum, respectively. The result shows a 0.8 % relative deviation between air and vacuum in 250- 300 nm and a wavelength dependent deviation, among which the maximum is a little more than 15%, in 300- 400 nm. The individual uncertainty caused by calibration condition of SBUS in vacuum is 1.8 % smaller than that in air. Mter theoretical analysis and confirmatory testing, it is found that the spectral reflectivity of Al+ MgF2 coating of SBUS reflecting element would be different in air and in vacuum, which proves the necessity of SBUS calibration in vacuum condition.
出处 《光学学报》 EI CAS CSCD 北大核心 2012年第1期22-27,共6页 Acta Optica Sinica
基金 国家自然科学基金(41074126) 吉林省青年科研基金(20090118)资助课题
关键词 遥感 辐射定标 光谱辐照度定标 Al+MgF2膜 紫外臭氧垂直探测仪 remote sensing radiometric calibration spectral irradiance calibration Al+ MgF2 coating solar backscattered ultraviolet spectroradiometer
  • 相关文献

参考文献13

二级参考文献48

共引文献66

同被引文献167

引证文献17

二级引证文献57

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部