期刊文献+

采用液滴分析技术定量检测水中溶解氧的研究 被引量:3

Quantitative detection of dissolved oxygen in the water by droplet analysis technology
原文传递
导出
摘要 在原有光纤-电容传感器的基础上,引入了光纤溶解氧(DO)传感器,在保持液滴分析仪原有结构的情况下,可同时获得水质的定性和定量信息。基于液滴分析技术(DAT)对水中30种不同浓度的DO含量进行了检测。实验结果表明,基于DTA,采用光纤DO传感器检测液体的DO含量是有效的。在考虑温度影响时,二次多项式拟合模型效果最佳,系统的灵敏度可达0.1mg/L,最大相对误差为2.38%。 Based on the original optical fiber-capacitance sensor,we introduce the dissolved oxygen(DO) optical fiber sensor,which can obtain the qualitative and quantitative information of water quality simultaneously with the structure of the droplet analyzer.Based on liquid drops analysis technique(DAT),30 different concentrations of DO have been tested.Experimental results show that it is feasible to test dissolved oxygen concentration with the proposed sensor using DAT.Considering the temperature affection,the quadratic polynomial model gives the best fitting result.The system′s sensitivity is 0.1 mg/L,and the maximal relative error is 2.38%.
出处 《光电子.激光》 EI CAS CSCD 北大核心 2012年第1期130-136,共7页 Journal of Optoelectronics·Laser
基金 国家自然科学基金(51005165/E051102)资助项目
关键词 液滴分析技术(DAT) 液滴指纹图 溶解氧(DO) 水质指标 定量检测 droplet analysis technology(DAT) liquid drop fingerprint dissolved oxygen(DO) water quality indicators quantitative detection
  • 相关文献

参考文献17

  • 1McMillan N D,Finlayson O,Fortune F,et al. The fiber drop analyser:a new multi-analyzer analytical instrument with applications in sugar processing and for the analysis of pure liquids[J]. Measurement Science & Technology, 1992,3(8) : 746-764.
  • 2宋晴,张国雄,裘祖荣.基于液滴指纹图波形分析的液体识别方法[J].光电工程,2005,32(4):19-23. 被引量:9
  • 3宋晴,张国雄,裘祖荣,张爱萍,史庆伟.光纤、电容液滴分析仪(二):系统设计[J].仪器仪表学报,2003,24(4):352-355. 被引量:4
  • 4Wang C H, Augousti A T, Mason J, et al. The capacitive drop tensibmeter-a novel multianalysing technique for measuring the properties of iiquids[J]. Measurement Science & Technology, 1999,10(1) : 19-24.
  • 5袁东,付大友,张新申,邹玉权,涂杰,李翔.溶解氧的测定方法研究进展[J].皮革科学与工程,2006,16(3):42-46. 被引量:16
  • 6XIONG Yan,XU Jing, ZHU Dao-qian, et al. Fiber-optic fluorescence sensor for dissolved oxygen detection based on fluorinated xerogel immobilized with ruthenium (II) complex[J].Journal of Sol-Gel Science and Technology, 2010,53(2) : 441-447.
  • 7WU Xiu-dong, CONG Yan, LIU Yan-hong, et al. Luminescence and oxygen sensing properties of ORMOSILs covalently grafted with a novel ruthenium (II) complex[J]. Journal of Sol-Gel Science and Technology,2009,49(3) : 355-363.
  • 8张建标,陈兴,黄俊,姜德生.一种可用于溶解氧测定的光纤氧传感器[J].传感器技术,2002,21(10):4-7. 被引量:21
  • 9Hsieh P T,Chen Y C,Lee M S,et al. The effects of oxygen concentration on ultraviolet luminescence of ZnO films by sol-gel technology and annealing[J]. Journal of Sol-Gel Science and Technology, 2008,47 ( 1 ) : 1-6.
  • 10Choi M M F,Xiao D. Single standard calibration for an optical oxygen sensor based on luminescence quenching of a ruthenium complex[J].Analytica Chimica Acta, 2000,403 (1-2) :57-65.

二级参考文献75

共引文献53

同被引文献31

  • 1吴英才,袁一方.基于折射率变化的一种新型盐度测量原理研究[J].光学学报,2005,25(2):199-202. 被引量:19
  • 2邢菁.计量标准装置测量重复性和稳定性的测试[J].计量与测试技术,2005,32(10):14-15. 被引量:11
  • 3McMillan N D, Finlayson O, Fortune F, et al. The fiber drop analyser: a new multianalyser analytical instrument with applications in sugar processing and for the analysis of pure liquids [J]. Measurement Science and Technology(S0957-0233), 1992, 3(8): 746-764.
  • 4McMillan N D, Smith S R P, Bertho A C, et al. Quantitative drop spectroscopy using the drop analyser: theoretical and experimental approach for microvolume applications of non-turbid solutions [J]. Measurement Science and Technology(S0957-0233), 2008, 19(5): 055601.
  • 5AugoustiAT, Mason J, McMillanND, etal. Application of capacitive tensiography to investigation of pendant drop growth [J]. Measurement Science and Technology(S0957-0233), 2006, 17(10): 48-52.
  • 6Smith S R P, O' Neill M, McMillan N D, et al. Testing by photometric measurement and camera study of theoretical prediction of microvolume universal sessile dropshape [C]//16th Conference in the Biennial Sensors and their Applications, Cork, Ireland, Sept l2-14, 2011, 307(1): 012013.
  • 7McMillan N, O'Neill M, Smith S, et al. Reliable TLDA-microvolume UV spectroscopy with applications in chemistry and biosciences for microlitre analysis and rapid pipette calibration [J]. Journal of Optics A: Pure and Applied Optics(S1464-4258), 2009, 11(5): 054016.
  • 8O'Neill M, McMillan N D, Smith S R P, et al. Performance Studies on the Transmitted Light Drop Analyser [C]// 16th Conference in the Biennial Sensors and their Applications, Cork, Ireland, Sept 12-14, 2011, 307(1): 012035.
  • 9方积乾.卫生统计学[M].6版.北京:人民卫生出版社,2012:156-166.
  • 10中华人民共和国卫生部,中国国家标准化管理委员会.GB/T57506-2006生活饮用水标准检验方法 金属指标[S].2006.

引证文献3

二级引证文献10

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部