摘要 目的探讨多台全自动血细胞分析仪测试结果的可比性和相关性。方法分析40份我院门诊及体检患者的新鲜血,标本以EDTA-K2作为抗凝剂,分别在LH750、MAXA、KX-21N(门)、KX-21N(急)血细胞分析仪上进行检测,对检测结果进行分析评估。结果各台仪器白细胞、红细胞、血红蛋白、红细胞压积、血小板5个项目结果密切相关,回归统计的斜率和截距可靠,偏差均在允许范围,检测结果具有可比性。结论4台仪器性能稳定,相关性良好,结果具有可比性,在临床中可同时或交替使用。
作者 孙颖
机构地区 高淳县人民医院
出处 《基层医学论坛》 2012年第1期86-87,共2页 The Medical Forum
  • 相关文献

参考文献3

二级参考文献9

共引文献29

同被引文献30

  • 1National Committee for Clinical Laboratory StandardS. Method comparison and bias estimation using patient samples approved guideline[S]. EP9-A,NCCLS, 1995.
  • 2Takubo T,Tatsumi N,Satoh N,et al,Evaluation of hematological values obtained with reference automated hematolog analyzers of six manufacters[J]. Southeast Asian J Trop Med Public Health, 2002,33(2) :62-67.
  • 3Kang SH,Kim HK,Ham CK,et al.Comparison of four hematologu analyzers, CELL-DYN Sapphire,ADVIA 120,Coulter LH 750,and Sysmex XE-2100,in terms of clinical usefulness[J].IntJ Lab Hematol,2008,30(6):480-486.
  • 4CarrJ, Geesaman S, Czader M. Performance evaluation of the new UniCel DxH800 coulter cellular analysis system in a large hospital setting[J]. Lab Medicine, 2012,43(5): 157-163.
  • 5Buttarello M, Plebani M. Automated blood cell counts: state of the art[J]. AmJ Clin Pathol , 2008 , 130(1) : 104-116.
  • 6Gulati G, SongJ, Florea AD, et al. Purpose and criteria for blood smear scan, blood smear examination, and blood sear review[J]. Ann Lab Med,2013, 33(1): 1-7.
  • 7Rao LV, Ekberg BA, Connor D, et al. Evaluation of a new point of care automated complete blood count(CBC)analyzer in various clinical settings[J]. Clin Chern Acta,2008,389(1-2):120-125.
  • 8Grecu DS,Paulescu E.Quality in post-analytical phase: indirect reference intervals for erythrocyte parameters of neonates[J]. Clin Biochem, 2013, 46(7-8) :617-621.
  • 9Kimbi HK,Sumbele IU , Nweboh M,et aI.Malaria and haematologic parameters of pupils at different altitudes aIong the slope of Mount Cameroon: a cross-sectional study[J].MalarJ ,2013,12: 193.
  • 10Wang H,LiuJ,Zhou Y,et al.Use of RBC-O and S-MCV parameters of SYSMEX XE-2100 in a patient with RBC cold agglutination[J]. Clin Lab,2013,59(1-2):217-219.