热固耦合式超声骨强度仪的可重复性及相关性研究 被引量:3

Study on the repeatability and correlation of warm-air and solid coupled mode ultrasound bone strength device
下载PDF
导出
摘要 目的:评估一款国产热固耦合式超声骨强度仪的可重复性和有效性。方法:选择11名志愿者,用该仪器测得右跟骨的两个超声参数,分别是超声速度(SOS)和宽带超声衰减(BUA),计算每个参数的变异系数均方根百分比(CVRMS%),评估该仪器的短期可重复性;选择18名志愿者,用类似的方法评估该仪器的中期可重复性;选择29名志愿者,用双能X线吸收法(DEXA)测量腰椎骨密度(BMD),并与用该仪器测得的跟骨SOS,BUA值进行相关性分析,评估该仪器的有效性。结果:①超声参数SOS,BUA的短期CVRMS%分别为0.30,4.24;中期CVRMS%分别为0.41,4.20。②SOS与腰椎BMD的相关系数r=0.486(P<0.01),BUA与腰椎BMD的相关系数r=0.629(P<0.01)。结论:与国际主流定量超声(QUS)检测仪器相比,热固耦合式超声骨强度仪具有较高的可重复性和准确性,可用于临床骨质状况评估。 Objective: To evaluate the repeatability and correlation of warm-air and solid coupled mode ultrasound bone strength device made in China. Method: Eleven and eighteen volunteers' right calcaneus ultrasound parameters, speed of sound (SOS) and broad- band ultrasound attenuation (BUA), were detected with the device, percentages of root-mean square of coefficient of variation (CVRMs%) of ultrasound parameters were calculated respectively, and used to assess the short- and middle-term repeatability of the device; 29 volunteers' right calcaneus ultrasound parameters (SOS, BUA) were de- tected with the device and bone mineral density (BMD) of lumbar spine were detected with dual energy X-ray absorptiometry (DEXA). Result: (~)The short-term CVRMs% were 4.24% for BUA, and 0.30% for SOS; the middle-term CVaMs% were 4.20% for BUA, and 0.41% for SOS. (~)The correlations between BMD of lumbar spine were 0.629 (P〈 0.01) for BUA and 0.486 (P 〈 0.01) for SOS. Conclusion: Compared to international top-grade quantitative ultrasound(QUS) testing device, the warm-air and solid coupled mode ultrasound bone strength device has high performance of repeatability and accuracy, so it can be used to assessing bone mass status. Author's address Affiliated Provincial Hospital of Anhui Medical University, Hefei, 230000
出处 《中国康复医学杂志》 CAS CSCD 北大核心 2012年第2期125-129,共5页 Chinese Journal of Rehabilitation Medicine
基金 中国科学院知识创新工程领域前言重点项目(0923A11291)
关键词 骨强度 定量超声 可重复性 相关性 双能X线吸收法 超声速度 宽带超声衰减 bone strength quantitative ultrasound repeatability correlation dual energy X-ray absorptiometry speed of sound broadband ultrasound attenuation
  • 相关文献

参考文献25

  • 1Riggs BL,Melton LJ 3rd.The worldwide problem of osteoporosis:insights afforded by epidemiology[J].Bone,1995,17(5Suppl):505S-511S.
  • 2Mazess R,Collick B,Trempe J,et al.Performance evaluation of a dual-energy x-ray bone densitometer[J].Calcif Tissue Int,1988,44(3):228-232.
  • 3Hans D,Fuerst T,Duboeuf F.Quantitative ultrasound bone measurement[J].Eur Radiol,1997,7(10):43.
  • 4Hans D,Schott AM,Meunier PJ.Ultrasonic assessment of bone:a review[J].Eur J Med,1993,2(3):157-163.
  • 5Hans D,Arlot ME,Schott AM,et al.Do ultrasound measurements on the os calcis reflect more the bone microarchitecture than the bone mass?:a two-dimensional histomorphometric study[J].Bone,1995,16(3):295-300.
  • 6Njeh CF,Boivin CM,Langton CM.The role of ultrosound in the assessment of osteoporosis:a review[J].Osteoporos Int,1997,7(1):7.
  • 7He YQ,Fan B,Hans D,et al.Assessment of a new quantitative ultrasound calcaneus measurement:precision and discrimination of hip fractures in elderly women compared with dual X-ray absorptiometry[J].Osteoporos Int,2000,11(4):354-360.
  • 8Njeh CF,Hans D,Li J,et al.Comparison of six calcaneal quantitative ultrasound devices:precision and hip fracture discrimination[J].Osteoporos Int,2000,11(12):1051-1062.
  • 9Stewart A,Reid DM.Precision of quantitative ultrasound:comparison of three commercial scanners[J].Bone,2000,27(1):139-143.
  • 10Lees B,Stevenson JC.Preliminary evaluation of a new ultrasound bone densitometer[J].Calcif Tissue Int,1993,53(3):149-152.

同被引文献63

引证文献3

二级引证文献9

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部