摘要 目的:通过对TECAN加样器检定结果不确定度的评定,初步探讨血站实验室对测量不确定度的评定方法。方法:依据中国合格评定国家认可委员会(CNAS)《测量不确定度要求的实施指南》及JJF1059-1999《测量不确定度评定与表示》等其他不确定度指南文件,采用A、B类评定方法对加样器加样量10μl测量不确定度进行评定。结果:加样器加样量10μl时扩展不确定度为U=(10±0.16)μl(K取2,置信区间95%)。结论:本实验所使用的方法可以对TECAN检定结果测量不确定度进行评定。 Objective:To preliminarily study the uncertainty evaluation method for blood center lab through the results from the TECAN automatic sampler. Method: All methods were used under the guidance of the China National Accreditation Service for conformity assessment document ((Guidance on the application of the requirements for measurement uncertainty)),]JF1059-1999((Guide to the expression of uncertainty in measurement)) and other re- lated regulation for uncertainty measurement. Type A and B assessment method were applied to measure the un- certainty during 10μl sample addition using the sampler. Result: The expended uncertainty of the sampler for 10μl sample addition was U = ( 10 ± 0. 1610) μl(K = 2, and the confidence interval was 95 %). Conclusion: The method used here could be applied to evaluate the uncertainty of the results from the TECAN sampler.
出处 《临床血液学杂志(输血与检验)》 2012年第1期86-87,共2页 Journal of Clinical Hematology(Blood Transfusion & Laboratory Medicine)
关键词 测量不确定度 加样器 uncertainty measurement TECAN sampler
  • 相关文献

参考文献2

二级参考文献9

  • 1Guldeline to Expression of Uncertainty in Measurement. ISO/TAG 4/WG3, June 1993.
  • 2Quantifying Uncertatnty in analytical chemistry. Vision 6, Eurachem,March 1995.
  • 3The expression of Uncertainty in Quantltative Testing. EAL23 August 1996.
  • 4ISO/TAG 4/WG3:Guide to the expression of uncertainty in measurement,June 1992.
  • 5ISO/IEC 17025:1999 General Requirements for the Competence of Calibration and Testing Laboratories,ISO,Geneva(1999).
  • 6ISO/DIS 15189.Quality Management in Medical Laboratories,ISO,Geneva (2000).
  • 7Ricos C,Alvarez V,et al.Current databases on biological variation:pros,cons and progress[J].Scand J Clin Lab Invest,1999,59 (11):491-500.
  • 8Fuentes-Arderiu X,Acebes-Frieyro G,Gavaso-Navarro L,et al.Premethrological (pre-analytical) variation of some biochemical quantities[J].Clin Chem Lab Med,1999,37(4):987-89.
  • 9吴雪祥.标准不确定度在卫生检验中的应用[J].中国公共卫生,2000,16(12):1142-1143. 被引量:1

共引文献57

同被引文献7

引证文献3

二级引证文献8

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部