摘要
目的:通过对TECAN加样器检定结果不确定度的评定,初步探讨血站实验室对测量不确定度的评定方法。方法:依据中国合格评定国家认可委员会(CNAS)《测量不确定度要求的实施指南》及JJF1059-1999《测量不确定度评定与表示》等其他不确定度指南文件,采用A、B类评定方法对加样器加样量10μl测量不确定度进行评定。结果:加样器加样量10μl时扩展不确定度为U=(10±0.16)μl(K取2,置信区间95%)。结论:本实验所使用的方法可以对TECAN检定结果测量不确定度进行评定。
Objective:To preliminarily study the uncertainty evaluation method for blood center lab through the results from the TECAN automatic sampler. Method: All methods were used under the guidance of the China National Accreditation Service for conformity assessment document ((Guidance on the application of the requirements for measurement uncertainty)),]JF1059-1999((Guide to the expression of uncertainty in measurement)) and other re- lated regulation for uncertainty measurement. Type A and B assessment method were applied to measure the un- certainty during 10μl sample addition using the sampler. Result: The expended uncertainty of the sampler for 10μl sample addition was U = ( 10 ± 0. 1610) μl(K = 2, and the confidence interval was 95 %). Conclusion: The method used here could be applied to evaluate the uncertainty of the results from the TECAN sampler.
出处
《临床血液学杂志(输血与检验)》
2012年第1期86-87,共2页
Journal of Clinical Hematology(Blood Transfusion & Laboratory Medicine)
关键词
测量不确定度
加样器
uncertainty measurement
TECAN sampler