可调式牙冠形态试件厚度控制仪的研制 被引量:2

Development of Adjustable Sculpting Apparatus for Dental Crown Shape Specimen
下载PDF
导出
摘要 目的:为能够严格控制牙冠形态试件各层结构尺寸的准确、均匀一致,设计出精度高、使用方便的可调式牙冠形态试件厚度控制仪。方法:利用AUTO CAD软件分析构图,我们自行研制了可调式牙冠形态试件厚度控制仪,用于牙冠形试件的核基底层蜡型制作、饰面层堆瓷过程中的厚度调控。应用此仪器制作0.75、1.00、1.20 mm等3种厚度的蜡型试件,将完成的试件从咬合面垂直切开,千分尺随机测量试件咬合面与轴壁厚度。结果:蜡型厚度误差为0.01 mm,同组蜡型的咬合面及轴壁各测试点之间的数值无显著差异(P>0.05)。结论:该仪器设计科学合理,结果准确可靠,试件加工重复性好,咬合面厚度与轴壁厚度均与所设计的理论厚度相符且均匀一致,能够满足口腔牙冠修复材料实验的测试要求。 Objective To design adjustable sculpting apparatus for dental crown shape specimen with high sensitivity and convenience in order to exactly control the accurateness and uniformness of each layer structure size of dental crown shape specimen. Methods With AUTO CAD software analyse, the sculpting apparatus to control the thickness for wax pattern of core crown and veneer specimen was developed. Three groups wax type specimens with the thickness of 0.75,1.00 and 1.20mm were made respectively. The wax pattern was cut vertically from the occlusal surface, the thickness of the occlusal surfaces and axial wall could be measured randomly, Results The thickness measurement error of the wax pattern was 0.01 mm, there was no significant difference between the thickness of occlusal surfaces and axial wall in the same group(P〉0.05). Conclusion Design of the sculpting apparatus is rational and scientific, the results of measurement are accurate and reliable, specimen processing is repeatable, and the practical size of occlusal surface thickness and axial wall thickness are in accord with the theory. The precision of the sculpting apparatus satisfies experiment testing requirements of dental restoration material.
出处 《医疗卫生装备》 CAS 2012年第1期26-27,33,共3页 Chinese Medical Equipment Journal
基金 陕西省科技计划项目(2008K14)
关键词 牙冠 蜡型 厚度控制仪 dental crown wax pattern sculpting apparatus for thickness
  • 相关文献

参考文献11

  • 1Wakabayashi N,Anusavice K J. C rack Initiation Modes in bilayered alumina/porcelain disks as a function of core/veneer thickness ratio and supporting substrate stiffness[J]. J Dent Res,2000,79 (6) : 1 398- 1 404.
  • 2Wagner W C,Chu T M. Biaxial flexural strength and indentation fracture toughness of three new dental core ceramics[J]. J Prosthet Dent, 1996,76:140-144.
  • 3Harrington Z A, McDonald, Knowles J. An in vitro study to investi- gate the load at fracture of Procera AllCeram crowns with various thickness of ocelnsal veneer porcelain[J]. Int J Prosthodont, 2003,16 ( 1 ) :54-58.
  • 4Sundh A,Sjogren G. A comparison of fracture strength of yttrium- oxide- partially-stabilized zirconia ceramic crowns with varying core thickness, shapes and veneer ceramics [J]. J Oral Rehabil, 2004,31 (7) : 682-685.
  • 5刘伟才,郭凌云,王海波,吴珺华,苏俭生.厚度配比对2种牙科陶瓷疲劳行为的影响研究[J].华西口腔医学杂志,2009,27(1):92-95. 被引量:2
  • 6Fleming G J, Dickens M,Thomas L J, et al. The in vitro failure of all- ceramic crowns and the connector area of fixed partial dentures using bilayered ceramic specimens : the influence of core to dentin thick- ness ratio[J]. Dent Mater, 2006,22 (8) : 771-777.
  • 7Fleming G J, Dickens M,Thomas L J,et al. The effect of core : dentin thickness ratio on the bi-axial flexure strength and fracture mode and origin of bilayered dental ceramic composites [J]. Dent Mater, 2005,21(2) : 164-171.
  • 8Neiva G,Yaman P,Dennison J B,et al. Resistance to fracture of three all-ceramic systems[J]. J Esthet Dent, 1998,10(2) :61-66.
  • 9Philp G K, Brukl C E. Compressive strengths of conventional, twin foil, and all-ceramic crowns[J]. J Prosthet Dent, 1984,52:215-220.
  • 10I Correr Sobrinho L,Consani S,Knowles J C. Effect of convergence angle and luting agent on the fracture strength of in eeram crowns[J]. J Mat Sei : materials in medicine, 1999(10) :493-496.

二级参考文献11

  • 1Chai H, Lawn BIL Wutfiphan S. Fracture modes in brittle coatingswith large interlayer modulus mismatch[J]. J Mater Res, 1999, 14 (9) : 3805-3817.
  • 2Fischer-Cripps AC, Lawn BR, Pajares A, et al. Stress analysis of elastic-plastic contact damage in ceramic coating on metal substrates[J]. J Am Ceram Soc, 1996, 79(10):2619-2625.
  • 3Zhao H, Hu XZ, Bush MB, et al. Contact damage in porcelain/ pd-alloy hilayers[J]. J Mater Res, 2000, 15(3):676-682.
  • 4Jung YG, Wuttiphan S, Peterson IM, et al. Damage modes in dental layer structures[J]. J Dent Res, 1999, 78(4):887-897.
  • 5Rhee YW, Kim HW, Deng Y, et al. Contact-induced damage in ceramic coatings on compliant substrates: Fracture mechanics and design[J]. J Am Ceram Soc, 2001, 84(5): 1066-1072.
  • 6Deng Y, Lawn BR, Lloyd IK. Characterization of damage modes in dental ceramic bilayer structures[J]. J Biomed Mater Res, 2002, 63 (2) : 137-145.
  • 7Miranda P, Pajares A, Guiberteau F, et al. Designing damageresistant brittle-coating structures: Ⅰ. Bilayers[J]. Acta Mater, 2003, 51 (203) :4347-4356.
  • 8Lawn BR, Pajares A, Zhang Y, et al. Materials design in the performance of all-ceramic crowns[J]. Biomaterials, 2004, 25 (14): 2885-2892.
  • 9Lee CS, Kim DK. Rate effects in critical loads for radial cracking in ceramic coatings[J]. J Am Ceram Soc, 2002, 85 (8):2019- 2024.
  • 10Fernandes CP, Glantz PO, Svensson SA, et al. A novel sensor for bite force determinations[J]. Dent Mater, 2003, 19(2): 118- 126.

共引文献1

同被引文献21

引证文献2

二级引证文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部