摘要 目的:探讨螺旋断层调强放疗治疗宫颈癌时MVCT引导下的治疗摆位误差。方法:2009年1月至2010年12月对30例宫颈癌术后的患者采用螺旋断层调强放疗治疗,每次治疗前实行靶区MVCT扫描,根据治疗计划的剂量分布对患者体位进行手动调节,获得治疗误差及分布规律。结果:30例患者共行MVCT扫描752次,X、Y、Z轴移动均数分别为(1.9+1.61)、(5.73+1.31)、(0.61+1.06)mm。结论:每次治疗前通过MVCT获得分次间摆位误差并对其加以纠正,这对提高宫颈癌放疗精度有积极意义。 Objective To analyze inter-fractional setup errors during helical tomotherapy for cervix cancer using daily online MVCT IGRT technique. Methods From Jan.2009 to Dec.2010, 30 cervix cancer patients were treated by helical tomotberapy. Before treatment, all patients underwent MVCT scanning. MVCT images were registered with automatic KVCT plan and used to readjust the registration according to the dose distribution. The setup errors were obtained and analyzed. Results A total of 752 online MVCT scannings were performed for these 30 patients. The setup errors on X,Y,Z directions were (1.9+1.61)mm, (5.73+1.31)mm and (0.61 +1.06)mm, respectively. Conclusion Daily online pre-treatment IGRT correction can reduce inter-fractional setup errors. MVCT can improve the accuracy and precisions of cervix cancer radiotherapy treatment.
出处 《医疗卫生装备》 CAS 2012年第1期65-66,共2页 Chinese Medical Equipment Journal
关键词 宫颈癌 图像引导放疗 摆位误差 螺旋断层放疗 cervix cancer IGRT setup error tomotherapy
  • 相关文献

参考文献9

  • 1Schulze D,Liang J,Yan D,et al. Comparison various online IGRT strategies:The benefits of online treatment plan re-optimization [J]. Radiother Oncol, 2009, 90 (30) : 367-376.
  • 2Riis H L,Zimmermann S J. Elekta Precise Table characteristics of IGRT remote table positioning [J]. Acta Oncol, 2009,48 (2) : 267- 270.
  • 3于金明,袁双虎.图像引导放射治疗研究及其发展[J].中华肿瘤杂志,2006,28(2):81-83. 被引量:99
  • 4Beavis A W. Is tomotherapy the future of IMRT [J]. Br J Radiol, 2004,77(916) :285.
  • 5Zeidan 0 A,Langen K M,Meeks S L,et ol. Evaluation of im- age2guidancep rotocols in the treatment of head and neck cancers[J]. Int J Radiat OncolBiol Phvs, 2007,67 (3) :670.
  • 6Meeks S L,Harmon J F,Langen K M ,et al. Performance character- ization of megavoltage computed tomograpyhy imaging on a helical TomoTherapy unit[J]. Med Phys, 2005,32 (8) :.2 673-2 681.
  • 7Forrest L J,Mackie T R,Ruchala K,et al. The utility of megavoh- age computed TomoTberapy images from a helical TomoTherapy sys- tem for setup verification purpose [J]. Int Radiat Oncal Biol Phys, 2004,60(5) : 1 639-1 644.
  • 8Shah A P,Langen K M,Ruchala K J,et ol. Patient dose from megavoltage computed imaging [J]. Int J Radiat Oncol Biol Phys, 2008,70(5) : 1 579-1 587.
  • 9Wilder R B,Chittenden L,Mesa A V,et a/. A prospective study of inteffraction prostate motion in the prone vs. supine position[J]. Int J Rodiat Oncol B iol Phys, 2010,77 (1) :165-170.