应用模拟定位机测量食管癌调强放疗中的摆位误差 被引量:6

Evaluation of set-up errors for patients for esophageal carcinoma treated with Intensity modulated radiation therapy by simulator
下载PDF
导出
摘要 目的:利用模拟定位机匹配功能测量食管癌调强放疗的摆位误差。方法:通过对171例食管癌患者计划系统生成的数字重建放射图(DRR)和模拟机拍摄的正、侧位图像的骨性标记进行比较,测出患者在左右(X轴)、头脚(Y轴)、前后(Z轴)方向误差,并对误差大于0.3cm进行调整。结果:患者左右、头脚、前后误差分别为0.19±0.24cm、0.27±0.34cm、0.27±0.25cm,其中左右方向需要调整的患者有40例,占总体的23.4%,头脚方向需要调整的患者有52例,占30.4%,前后方向需要调整的患者有64例,占37.4%。结论:模拟机图像匹配功能的应用,能对患者摆位误差进行及时的调整,从而提高实际摆位精度,同时为正确设定计划靶体积(PTV)提供了依据。 ObjeetiveTo evaluate setup error by Varian Acuity simulator Methods 171 patients for esophageal carcinoma treated with Intensity modulated radiation therapy were included in this study. DRRs of each patient were compared with portal images from simu- lator and set - up errors were recorded and analyzed. Patients should be adjusted when setup error exceeded 0. 3 cm. Results The setup errors in Right - Left, Cranial - Caudal and Anterior - Posterior directions were (0. 19 +0. 24) cm, (0. 27 +0. 34) cm and (0. 27 + 0. 25 ) cm 40 patients should be adjusted in Right - Left direction and 52 patients in Cranial - Caudal direction , 64 patients in Anterior -Posterior direction. And the margins from CTV or ITV to PTV were given. Conclusion Measurement of setup error before radiation delivery using simulator with correction could greatly improve the patient setup precision and may guide optimal PTV designing.
出处 《医疗装备》 2012年第2期10-12,共3页 Medical Equipment
关键词 摆位误差 模拟定位机 调强放疗 功能测量 食管癌 应用 图像匹配 癌患者 esophageal carcinoma Intensity modulated radiation therapy Setup error
  • 相关文献

参考文献7

二级参考文献27

  • 1何正中,戴建荣,乐文友.调强放疗中摆位误差对脊髓剂量的影响[J].中华放射肿瘤学杂志,2007,16(2):132-135. 被引量:12
  • 2Bijhold J, van Herk M, Vijlbrief R, et al. Fast evaluation of patient set-up during radiotherapy by aligning features in portal and simulator images. Phys Med Biol, 1991, 36 : 1665-1679.
  • 3Remeijer P, Geerlof E, Ploeger L, et al. 3-D portal image analysis in clinical practice: an evaluation of 2-D and 3-D analysis techniques as applied to 30 prostate cancer patients. Int J Radiat Oncol Biol Phys, 2000, 46:1281-1290.
  • 4Fu W, Yang Y, Li X, et al. Dosimetric effects of patient rotation- al setup errors on prostate IMRT treatments. Phys Med Biol, 2006, 51:5321-5331.
  • 5Hurkmans CW, Remeijer P, Lebesque JV, et al. Set-up verification using portal imaging;review of current clinical practice. Radiother Oncol,2001,55 : 105-120.
  • 6van Asselen B, Dehnad H, Raaijmakers CP, et al. Implanted gold markers for position verification during irradiation of head-and- neck cancers:a feasibility study. Int J Radiat Oncol Biol Phys, 2004, 59 : 1011-1017.
  • 7Siebers JV, Keall PJ, Wu Q, et al. Effect of patient setup errors on simultaneously integrated boost head and neck IMRT treatment plans. Int J Radiat Oncol Biol Phys, 2005, 63:422-433.
  • 8Ranald I, Paul A, John P, et al.Patient positioning using detailed three- dimensional surface data for patients undergoing confounal radiation therapy for carcinoma of the prostate: A feasibility study. Int. J. Radiation Oncology Biol. Phys. 2001, 49:p225 230.
  • 9Gilbeau L, Octave- Prignot M, Loncol T, et al.Compurison of setup accuracy of three different thermoplastic masks for the treatment of brain and head and neck tumors. Radiotherapy and Oncology .2001, 58: 155-162.
  • 10Kaiser A, Schultheiss TE, Wong JY, et al. Pitch, roll, and yaw variations in patient positioning. Int J Radiat Oncol Biol Phys, 2006,66:949-955.

共引文献73

同被引文献39

  • 1汪志,唐虹,李锐,洪浩,孔令玲.利用模拟机对493例患者摆位误差的测量[J].中国医学物理学杂志,2011,28(4):2733-2735. 被引量:4
  • 2林承光,邓小武,吴裕起,黄峻,苏建新,伍建华,何晓华.放射治疗实施过程中的质量保证及质量控制[J].中国肿瘤,2004,13(9):544-545. 被引量:18
  • 3戴建荣,胡逸民.图像引导放疗的实现方式[J].中华放射肿瘤学杂志,2006,15(2):132-135. 被引量:184
  • 4于龙珍,翟振宇,沈君姝.体部肿瘤精确放疗摆位误差分析[J].中国医学物理学杂志,2007,24(3):160-162. 被引量:36
  • 5Xing L,Thorndyke B,Schreibmann E,et al. Overview of image-guided radiation therapy [J]. Medical Dosimetry, 2006,31(2) :91-112.
  • 6Cox JD,Stetz J, Pajak TF. Toxicity criteria of the Radia- tion Therapy Oncology Group (RTOG) and the European Organization for Research and Treatment of Cancer (E- ORTC) [J]. Int J Radiat Oncol Biol PhyS, 1995,31 : 1341- 1346.
  • 7Emiliano S,Patrick D,Richard J,et al. Patient--specific three- dimensional concomitant dose from cone beam computed tomography exposure in image-guided radiotherapy [J]. lnt J Radiat Oneol Biol Phys, 2012,83 ( 1 ) : 419-426.
  • 8Rietzel E,Pan T, Chen GT, et al. Four-dimensional com- puted tomography:image formation and clinical protocol [J]. Med Phys, 2005,32(4) : 874-889.
  • 9Mackie TR, Kapatoes J, Ruchala K, et al. Image guidance for percise conformal radiotherapy [J]. Int J Radiat Oncol Biol Phys, 2003,56 ( 1 ) : 89-105.
  • 10殷旭东,陈广涛,翟振宇,李建中,王静,汪琪.利用模拟机对鼻咽癌调强放疗进行位置验证[J].中国医学物理学杂志,2008,25(2):560-563. 被引量:10

引证文献6

二级引证文献25

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部