摘要 目的 回顾性分析金属置入物的老年患者接受3T MRI检查的安全性,并评估其图像质量.方法 345例金属置人物的老年患者接受3T MRI检查,置人物为非铁磁性或弱磁性金属.观察检查前和检查后即刻、1W 患者生命体征和不适症状,并评估图像质量.结果 所有患者均未出现由检查引发的不良反应,3个时间段观测的患者生理指标稳定,组间差异无统计学意义(P〉0.05).图像符合诊断要求.结论非铁磁性或弱磁性的体内金属置入物患者接受3T MRI检查安全可行,未明显影响图像诊断效果. Objective To validate the safety of metal implants in 3T MRI examination in patients via the cases follow- up and observation and to assess the image quality. Methods 345 elderly cases with medical metal implants were included in the study. Physical indexes and body complaints (if any) were recorded in three phase: before, soon after and 1 week after the MRI examination. Image quality and statistical analysis were performed. Results No side effect from the MRI examination was recorded. Physical indexes of patients obtained in the three phases were stabilized and showed no statistical difference between groups (P〉 0.05). Image quality satisfied diagnostic requirement. Conclusions 3T MRI examination in the patients with metal implants is safe on condition that the implants are non-ferromagnetic or weakly ferromagnetic. Images quality are not obviously influenced for diagnosis.
机构地区 [
出处 《老年医学与保健》 CAS 2012年第1期48-51,共4页 Geriatrics & Health Care
关键词 金属置人物 老年患者 3T MRI 安全性 Metal implant Elderly patient 3TMRI Safety
  • 相关文献

参考文献8

  • 1Shafiei F,Honda E,Takahashi H,et al.Artifacts from dental casting alloys in magnetic resonance imaging[J].J DentRes,2003,82 (8):602-606.
  • 2嵇鸣,叶春涛,苗华栋,林光武,臧雪如,朱震方.金属置入物患者行3.0T MR检查的安全性探讨[J].中国医学计算机成像杂志,2009,15(6):570-573. 被引量:7
  • 3Shellock FG.Biomedical implants and devices:assement of mastic field interactions with a 3.OTesla MR system[J].J Magn Reson Imaging,2002,16 (6):721-732.
  • 4Shellock FG.Magnetic resonance safety update 2002:implants and devices[J].J Magn Reson Imaging,2002,16 (5):485-496.
  • 5Spuentrup E,Ruebben A,Schaeffter T,et al.Magnetic resonance-guide coronary artery stent placement in a Swie mod-el[J].Circulation,2002,105 (7):874-879.
  • 6Hug J,Nagel E,Bornstedt A.et al.Coronary arterial stents:safety and artifacts during MR imaging[1].Radiology,2000,216 (3):781-787.
  • 7贺李,余深平,孟悛非.金属物对磁共振检查的影响及相关研究进展[J].国际医学放射学杂志,2009,32(1):53-57. 被引量:10
  • 8Hahnel S,Nguyen TH,Rohde S,et al.3.OTesla contrast-enhanced MR angiography of carotid artery stents:in vitro me-asurements as compared with 1.5Tesla[J].J Neuroradiol,2006,33 (2):75-80.

二级参考文献26

  • 1靳二虎,马大庆.MR检查时人体内生物医学置入物和置入设备的安全性研究进展及对策[J].中华放射学杂志,2004,38(9):999-1001. 被引量:11
  • 2陈宝峰,徐新.起搏器及植入型心律转复除颤器患者磁共振成像检查的安全性及相容性[J].中华心律失常学杂志,2007,11(1):75-78. 被引量:7
  • 3Capelastegui A,Fernandez-Canton G,Fernamdez-Ruanova B,et al.The safety of magnetic resonance imaging:an analysis based on a review of incidents at the Osatek Center[J].Badiologia,2006,48:225-234.
  • 4Hennemeyer CT,Wicklow K,Feinberg DA,et al.In vitro evaluation of platinum Guglielmi detachable coils at 3 T with a porcine model:safety issues and artifacts[J].Radiology,2001,219:732-737.
  • 5Shafiei F,Honda E,Takabashi H,et al.Artifacts from dental casting alloys in magnetic resonance imaging[J].J Dent Res,2003,82:602-606.
  • 6Weishaupt D,Quick HH,Nanz D,et al.Ligating dips for three-dimensional MR angiography at 1.5 T:in vitro evaluation[J].Radiology,2000,214:902-907.
  • 7Hahnel S,Nguyen-Trong TH,Rohde S,et al.3.0 Tesla contrastenhanced MR angiography of carotid artery stents:in vitro measurements as compared with 1.5 Tesla[J].J Neuroradiol,2006,33:75-80.
  • 8Matsuura H,Inoue T,Ogasawara K.et al.Quantitative analysis ofmagnetic resonance imaging susceptibility artifacts caused by neurosurgical biomaterials:comparison of 0.5,1.5,and 3.0 Tesla magnetic fields[J].Neurol Med Chit (Tokyo),2005,45:395-398.
  • 9Destine D,Mizutani H,Igarashi Y,et al.Metallic artifacts in MRI caused by dental alloys and magnetic keeper[J].Nihon Hotetsu Shika Gakkai Zasshi,2008,52:205-210.
  • 10Starcukova J,Starcuk Z Jr,Hubakova H,et al.Magnetic susceptibility and electrical conductivity of metallic dental materials and their impact on MR imaging artifacts[J].Dent Mater,2008,24:715-723.

共引文献13

同被引文献22

引证文献5

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部