7600-020型和Vitros-350型生化分析仪测定血清电解质结果的比对分析 被引量:6

The comparison of serum electrolyte results by 7600-020 and Vitros-350 biochemical analyzer
下载PDF
导出
摘要 目的:比对分析日立7600-020型全自动生化分析仪与美国强生Vitros-350型干式化学分析仪测定血清电解质的结果,并作偏差评估,以保证测定结果的一致性。方法:按照美国临床实验室标准化委员会《EP9-A2文件》的要求,以7600-020系统为比对方法,Vitros-350系统为试验方法,分别在2台仪器上测定血清电解质,并记录数据,检查离散点,计算线性方程和相关系数(r),并对其进行偏倚评估。结果:2台仪器3个项目的检测结果均具有较好的相关性(P<0.05),临床接受性评价均未超过CLIA'88允许误差的一半。结论:日立7600-020型全自动生化仪和强生Vitros-350型干式生化分析仪电解质测定结果基本一致,具有很好的可比性,临床检测过程中可任选一种进行电解质测定。 Objective:To evaluate the results of serum electrolytes by two automatic biochemistry analyzer systems:7600-020 full-automatic biochemical analyzer(Hitachi,Japan) and VITROS 350 dry-type biochemistry analyzer(Johnson,US),and assess the bias,which to ensure the consistency of test results.Methods:According to the National Committee for clinical laboratory standards EP9-A documents,7600-020 system was as control,Vitros-350 system was as experimental instrument,measure serum electrolytes by the two instruments,record the data,check the discrete point,calculate linear equations and correlation coefficients(r),and carry on the bias estimation.Results:Correlations between the two instruments were fine enough(P0.05) and the clinical acceptability assessments did not exceed half of the acceptable error by CLIA′88.Conclusions:The electrolyte results which determined by Hitachi 7600-020 automatic biochemical analyzer and Johnson Vitros-350 dry type biochemistry analyzer are basically consistent and have a good comparability,either of the two instruments can be selected to do electrolyte determination.
作者 孙丽
出处 《蚌埠医学院学报》 CAS 2011年第12期1385-1387,共3页 Journal of Bengbu Medical College
关键词 电解质 检测系统 比对试验 electrolytes detection system control experiment
  • 相关文献

参考文献6

二级参考文献17

共引文献24

同被引文献35

引证文献6

二级引证文献24

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部