成人腹部CT图像质量与体型参数相关性的探讨 被引量:3

Research about Relationship of Parameters of Body Size and Image Quality in Abdominal CT of Adults
下载PDF
导出
摘要 目的考察各种体型参数与腹部CT图像质量的关系。方法选取符合标准的50名成人进行腹部CT检查,设计不同管电流-时间乘积(mAs)扫描(50、100、150、200、250、300mAs),规范其它扫描参数和图像滤波条件,同时记录患者体重、身高、腹围数据,测量CT图像L1椎弓根水平横断面的面积、前后径、左右径和周径数值,考察各种体型参数与图像质量之间的关系。结果各体型参数与图像噪声的相关系数分别是:0.713(体重)、0.388(身高)、0.687(体重身高比)、0.685(体表面积)、0.767(横断面左右径)、0.803(横断面面积)、0.765(横断面前后径)、0.739(横断面周径)。实际腹围测量与横断面周径之间比较无显著统计学差异(P=0.412)。结论体型参数与图像噪声的相关性很高,腹围可以作为调节成人腹部CT扫描参数的选择。 Objective To investigate the relationships between several body size parameters and the image quality.Methods 50 adults who performed abdominal CT were selected,received scanning with different tube current-time product setting(50,100,150,200,250,300) but fixing another exposure parameters and condition of image filter.At the same time,weight,height,abdomen circumference were recoded;area,sagittal and coronal diameter,circumference of the cross-section of the L1 lateral neck of vertebra were measured.Finally,the relationship between body size parameters and image quality were investigated,and extracted some parameters which were convenient and easy to manipulate and with a certain correlation.Results The correlation coefficient between the CT image quality and the weight was 0.713,the height was 0.388,the ratio of weight and height was 0.687,the body surface area 0.685,the coronal diameter 0.767,the area of the cross-section 0.803,the sagittal diameter 0.765,and the circumference 0.739,respectively.There were no statistical difference between the cross-section circumference in the image and the abdominal circumference(P=0.412).Conclusion The relationship of all the body size parameters and the image quality is very high,the circumference of abdomen can be used as a parameter adjusting the scanning parameters in abdominal CT adults.
出处 《中国CT和MRI杂志》 2012年第1期71-73,共3页 Chinese Journal of CT and MRI
关键词 体层摄影术 X线计算机 图像质量 腹部 体型 参数 Tomography X-Ray computed image quality abdomen body size parameter
  • 相关文献

参考文献15

  • 1Aldrich JE,Chang SD,Bilawich A-M,et al.Radiatin dose in abdominal computed tomography:the role of patient size and the selection of tube current[J].Can Assoc Radiol J 2006,57(3):152-158.
  • 2Prasa SR,Wittran C,Shepard JA,et al.Standard-dose and 50% reduced-dose chest CT:comparing the effect on image quality[J].AJR Am J Roentgenol,2002,179(2):461-465.
  • 3Haaga JR.Radiation dose management:weighing risk versus benefit(commentary)[J].AJR Am J Roentgenol,2001,177(2):289-291.
  • 4高德春,吴苔,张洪锡.儿童腹部CT曝光参数设置[J].中华放射医学与防护杂志,2004,24(4):362-363. 被引量:7
  • 5刘晓东,龚建平,王建玲,朱江涛,钱铭辉.患者体重参数与腹部CT图像质量的关系[J].实用放射学杂志,2007,23(8):1062-1064. 被引量:3
  • 6Cohen M,Fischer H,Hamacher J,et al.CT of the head by use of reduced current and kilovoltage:relationship between image quality and dose reduction[J].AJNR Am J Neuroradiol,2000,21(9):1654-1660.
  • 7Menken J.Comparison of different body size parameters for individual dose adaption in body CT of adults[J].Radiology,2005,236(2):565-571.
  • 8赵松山 刘友梅 等.中国成年男子体表面积的测量[J].营养学报,1984,6:87-95.
  • 9赵松山,刘友梅,姚家邦,杨增仁,梁月琴,张是敬.中国成年女子体表面积的测量[J].营养学报,1987,10(3):200-207. 被引量:41
  • 10Nakayama Y,Awai K,Funama Y,et al.Abdominal CT with low tube voltage:preliminary observations about radiation dose,contrast enhancement,image quality,and noise[J].Radiology,2005,237(3):945-951.

二级参考文献20

  • 1赵松山 刘友梅 等.中国成年男子体表面积的测量[J].营养学报,1984,6:87-95.
  • 2Huda W,Atherton JV,Ware DE,et al.An approach for the estimation of effective radiation dose at CT in pediatric patients.Radiology,1997,203:417-422.
  • 3Boone JM,Geraghty EM,Seibert JA.Dose reduction in pediatric CT:a rational approach[J].Radiology,2003,228(2):352-360.
  • 4Huda W,Scalzetti EM,Levin G.Technique factors and image quality as functions of patient weight at abdominal CT[J].Radiology,2000,217(2):430-435.
  • 5Prasad SR,Wittram C,Shepard JA,et al.Standard-dose and 50% reduced-dose chest CT:comparing the effect on image quality[J].AJR,2002,179(2):461-465.
  • 6Mahnken AH,Widberger JE,Simon J,et al.Detection of coronary calcifications:feasibility of dose reduction with a body weight-adapted examination protocol[J].AJR,2003,181(8):533-538.
  • 7Irie T,Inoue H.Individual modulation of the tube current-seconds to achieve similar levels of image noise in contrast-enhanced abdominal CT[J].AJR,2005,184(5):1514-1518.
  • 8Kalra MK,Prasad S,Sain S,et al.Clinical comparison of standard-dose and 50% reduced-dose abdominal CT:effect on image quality[J].AJR,2002,179(5):1101-1106.
  • 9Kalra MK,Maher MM,Toth TL,et al.Strategies for CT radiation dose optimization[J].Radiology,2004,230(3):619-628.
  • 10Haaga JR.Radiation dose management:weighing risk versus benefit (commentary)[J].AJR,2001,177(2):289-291.

共引文献52

同被引文献33

  • 1刘晓东,龚建平,王建玲,朱江涛,钱铭辉.患者体重参数与腹部CT图像质量的关系[J].实用放射学杂志,2007,23(8):1062-1064. 被引量:3
  • 2白人驹.医学影像诊断学[M].第2版,北京:人民卫生出版社,2009:403-619.
  • 3Aldrich JE, Chang SD, Bilawich AM, et al. Radiation dose in abdominal computed tomography., the role of patient size and the selection of tube current [J]. Can Assoc Radiol J, 2006, 57: 152-158.
  • 4Prasa SR, Wittran C, Shepard JA, et al. Standard-dose and 50% reduced-dose chest CT: comparing the effect on image quality [J]. AJR, 2002, 179: 461-465.
  • 5Irie T, Inoue H. Individual modulation of the tube current- seconds to achieve similar levels of image noise in contrast-en-hanced abdominal CT [J]. AJR, 2005, 184: 1541-1518.
  • 6Kalra MK, Maeher MM, Kamath RS, et al. Sixteen-detector row CT of abdomen and pelvis; study for optimization of Z- axis modulation technique performed in 153 patients [J]. Ra- diology, 2004, 233 : 241-249.
  • 7Nakayama Y, Awai K, Funama Y, et al. Abdominal CT with low tube voltage: preliminary observations ahout radia- tion dose, contrast enhancement, imagequality, and noise [J]. Radiology, 2005, 237: 945-951.
  • 8Das M, Mahnken AH, Muhlenbruch G, et al. Individual a- dapted examination protocol for reduction of radiation expo- sure for 16 MDCT chest examinationa [J]. AJR, 2005, 184: 1437-1443.
  • 9Haaga JR. Radiation dose management:weighting risk versus benefit (commentary) [J]. AJR, 2001, 177: 289-291.
  • 10Suess C, Chen X. Dose optimization in pediatric CT: current technology and future innovations [J]. Pediatr Radiol, 2002, 32: 729-734.

引证文献3

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部