暴露于空气不同时间对血型试剂性能的影响研究 被引量:1

The research of the influence of BG reagent's capability when exposed in the air in different period of time
下载PDF
导出
摘要 目的:正确鉴定红细胞血型是确保输血安全的首要条件,但在临床输血和采供血机构血型判定等实际工作中,会出现ABO血型漏检或假阳性。对血型试剂性能的评价及其影响因素的研究可以更好地帮助临床正确鉴定血型。方法:本实验将通过对两种国产ABO血型试剂及Rh(D)血型试剂的效价及亲和力在暴露在空气不同时间条件下的变化来进行比较,从而观察在这种条件下对血型鉴定结果有无影响。结果:试剂在暴露在空气(室温)不同时间后所测的效价变化不大,但暴露在空气(室温)不同时间对亲和力的影响随时间的延长而减弱,特别是Rh(D)血型试剂亲和力变化较大。结论:在检测血型时试剂不能暴露在空气太长时间,否则会导致亲和力降低而错判结果。 Objective:To exactly authenticates the red cell blood type is initial condition that insures the blood transfusion safety. But in the actual work which examine blood group, the organization of clinical transfuse and adopt provide the blood will occasionally make the wrong effect such as false negative or false positive. Try to evaluate the function of BG reagent and study its influential factor with help the clinic to type the BG perfectly. Methods:This experiment will examine the potency and affinity of two kinds of monoclonal anti -A (B)reagent and Rh (D) re- agent under the condition of expose in the air in defferent period of time, so that to observe if there would be influence the result in typing BG ex- amine on that condition. Results:There are almost no effection to the potency of the BG reagent while expose it in the air( indoor temperature) in defferent period of time. However,it is have effection to the affinity of the BG reagent under that condition. It will be weaker and weaker as the time of exposing in the air extend. Conclusion:While taking BG examination,we can not expose the BG -reagent too long in the air, otherwhise, the affinity will step down and lead to the error result.
作者 梁剑晖
出处 《中外医学研究》 2012年第7期4-6,共3页 CHINESE AND FOREIGN MEDICAL RESEARCH
关键词 ABO血型 效价 亲和力 单克隆试剂 ABO BG(blood group) Potency Affinity Mab - reagent
  • 相关文献

参考文献9

  • 1孙爱农,吴晓燕.单克隆试剂误诊ABO血型初探[J].中国输血杂志,1999,12(3):157-158. 被引量:16
  • 2孙爱农,邓星河,吴晓燕.单克隆试剂导致ABO定型假阳性结果的观察[J].广东医学,2000,21(12):1053-1054. 被引量:4
  • 3杨毓明,陈卫红,邓曦,万丽萍,纪红.单克隆抗体试剂漏检ABO血型2例[J].中国输血杂志,2001,14(2):104-105. 被引量:15
  • 4Beck ML,Kowalski MA,Kirkegaard J,et al.Unexpected activitywith mnoclonal anti-B reagents[J].Immunohematology,1992,8(1):22-26.
  • 5Garratty G,Arndt P,Co A,et al.Fatal hemlytic transfusion reac-tion resulting from ABO mlstyping of a patient with acquired B antigen detectable only by some mnoclonal anti-B reagents[J].Transfusion,1996,36(3):351-355.
  • 6Walker RH,Branch DR,Dzik WH,et al.Technial Manual Llthed[J].Bethesda:AABB,1993,22(3):170-173,208-210.
  • 7李勇.人ABO血型单克隆抗体试剂的特性及质量要求[J].中国生物制品学杂志,1994,7(1):47-47.
  • 8孙爱农,冯志广,李兆兰,何锐洪,吴志峰,廖艳婷.单克隆抗-A(B)试剂质量鉴定与分析[J].中国误诊学杂志,2004,4(1):34-36. 被引量:8
  • 9中国生物制品标准化委员会.抗A、抗B血型定型试剂(单克隆抗体)制造及检定规程[M].见:中国生物制品规程2000年版.北京:化学工业出版社,2000:601-604.

二级参考文献11

  • 1上海生物制品研究所血型组.血型与血库[M].上海:上海为民出版社,1997.26-28.
  • 2李勇.人ABO血型单克隆抗体试剂的特性及质量要求[J].中国生物制品学杂志,1994,7(1):47-47.
  • 3大久保康人 李慧文译.血型与输血检查[M].北京:中国科学技术出版社,1996.3-25,132-135.
  • 4Scott ML, Voak D .Detection of acquired B antigen by mnoclonal anti-B blood grouping reagents. Transfusion, 1997,37(1):103-104.
  • 5Butch SH,Judd WJ,Steiner EA,et al.Electronic verification of donor-recipient compatibility:the computer crossmatch. Transfusion,1994,34 (2):105.
  • 6Garratty G, Arndt P, Co A, et al.Fatal hemlytic transfusion reaction resulting from ABO mistyping of a patient with acquired B antigen detectable only by some mnoclonal anti-B reagents. Transfusion, 1996,36(3):351-355.
  • 7Beck ML, Kowalski MA, Kirkegaard J, et al. Unexpected activity with mnoclonal anti-B reagents. Immunohematology,1992, 8(1):22-26.
  • 8Cameron C,Graham F,Dunsford I,et al.Acquisition of a B-like antigen by red blood cells.Br Med J,1959,2(1):29-31.
  • 9孙爱农,吴晓燕.单克隆试剂误诊ABO血型初探[J].中国输血杂志,1999,12(3):157-158. 被引量:16
  • 10杨毓明,陈卫红,邓曦,万丽萍,纪红.单克隆抗体试剂漏检ABO血型2例[J].中国输血杂志,2001,14(2):104-105. 被引量:15

共引文献30

同被引文献4

  • 1周华友,兰炯采.人类Rh血型系统研究进展[J].中国输血杂志,2004,17(3):211-214. 被引量:38
  • 2医学实验室质量和能力认可准则,CNAS-CL02,5.3.2.3.
  • 3Hillyer CD, Shaz BH, Winkler AM, Reid M. integrating molecular technologies for red blood cell typing and compatibility testing into blood centers and transfusion services. Transfus Med Rev. 2008 Apr;22 (2) : 117 : 32. doi: 10. 1016/j. tmrv. 2007. 12. 002. PubMed PMID: 18353252.
  • 4Yu Y, Ma C, Feng Q, Chen X, et al. Establishment and perform- ance assessment of preparation technology of internal quality control products for blood transfusion compatibility testing. Exp Ther Med. 2013 May;5(5):1466 :1470. Epub2013 Mar7. PubMed PMID: 23737900 ; PubMed Central PMCID : PMC3671829.

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部