摘要 分析了国际计量局组织的137Cs空气比释动能关键比对(BIPMRI(I)-K5)基准的几何参数、修正因子以及修正因子的确定方法,对我国137Cs空气比释动能基准电离室的设计,拟相关修正因子测量所采用的物理实验方法和模拟计算等提供了有益的借鉴。 The analysis and comparison of the geometric parameters, the correction factors and the method to determine the correct factor of 137 Cs air kerma primary standard which participated the 137 Cs air kerma key comparison BIPM RI (I)-K5 ,organized by International Bureau of Weights and Measures (BIPM) , will provide useful reference about the design in ionization chamber, the measurement of correction factors and the method of simulation for our national primary standard of air kerma for 137 Cs gamma-rays.
出处 《计量学报》 CSCD 北大核心 2011年第B12期36-40,共5页 Acta Metrologica Sinica
关键词 计量学 空气比释动能 国际比对 电离室 修正因子 Metrology Air kerma International comparisons Ionization chamber Correction factors
  • 相关文献

参考文献29

  • 1Perroche A M, Boutillon M. Comparisons and calibrations at the BIPM ( 1993 - 1995) [C]//Document CCEMRI (I)/95 -2, submitted to the 12^th Meeting of the CCEMRI, 1995.
  • 2Loftus T P, Weaver J T. Standardization of ^60Co and ^137Cs gamma-ray beams in terms of exposure [ J ]. J Res Natl Bur Stand( Sect A), 1974, 78A :465 - 476.
  • 3Witzani J, Leitner A, Air kerma standard at the BEV for ^137Cs rays [ J ]. Radiat Prot Dosim, 1996, 67, 291 - 294.
  • 4Kessler C, Allosy-Roberts P J, Steurer A, et al. Comparison of the standards for air kerma of the BEV and the BIPM for ^137Cs gamma radiation [R]. BEV, 2010.
  • 5Allisy-Roberts P J, Boutillon M, et al. Comparisons of the standards for air kerma of the BEV and the BIPM for ^137Cs and ^60Coγrays[R]. BIPM, 1996.
  • 6Allisy-Roberts P J, Boutillon M. Comparison of the standards for air kerma of the BNM-LPRA and the BIPM for 137Cs γ rays [R]. BIPM, 1996.
  • 7Allisy-Roberts P J, Kessler C, Toni M, et al. Comparison of the standards for air kerma of the ENEA-INMRI and the BIPM for ^137Cs γ rays [R]. BIPM,2010.
  • 8Kessler C, Allisy-Roberts P J, Burns D. T. Comparison of the standards for air kerma of the ITN(Portugal) BIPM for ^137Cs γ rays[ R]. BIPM ,2008.
  • 9Kessler C, Roger P, Allisy-Roberts P J. Chul-Young Yi, Comparison of the standards for air kerma of the KRISS and the BIPM for ^137Cs gamma radiation [R]. BIPM, 2010.
  • 10Allisy-Roberts P J, Kessler C, Burns D T. Comparison of the standards for air kerma of the LNE-LNHB and the BIPM for ^137Cs gamma radiation[R]. BIPM,2008.

同被引文献10

  • 1胡家成,杨小元,杨元第,张辉,樊铁栓,孙吉宁,黄斐增,刘鉴常,叶卫国.球-圆柱形空腔电离室壁修正因子的实验测量[J].计量学报,2007,28(z1):256-260. 被引量:2
  • 2韩奎初.测量照射量(空气比释动能)的空腔电离室[J].核标准计量与质量,1995,0(1):33-35. 被引量:3
  • 3胡家成,杨小元,杨元第,樊铁栓,黄斐增,刘鉴常,孙吉宁,叶卫国.球形石墨空腔电离室壁修正因子的测量与验证[J].中华放射医学与防护杂志,2005,25(6):561-564. 被引量:5
  • 4阿蒂克斯FH,罗奇wc.辐射剂量学:第1卷[M].北京:原子能出版社,1981.
  • 5Kessler C, Roger P, Allisy-Roberts P J, et al. Comparison of the standards for air kerma of the KRISS and the BIPM for mCs gamma radiation[ J]. Metrologia, 2010, 47(S) : 06022.
  • 6Allisy-Roberts P J, Kessler C, Toni M, et al. Comparison of the standards for air kerma of the ENEA- INMRI and the BIPM for 137Cs'yrays[ R]. Rapport BIPM- 2010/7,2010.
  • 7Allisy-Roberts P J, Bums D T, Takate N, et al. Comparison of the standards for air kerma of the NMIJ and the BIPM for 137Cs '~ rays[ R]. Rapport BIPM-2004/12, 2004.
  • 8Rogers D W O, Treumiet J. Monte Carlo calculated wall.and axial non-uniformit corrections for primary standards of air kerma[ R]. NRCC Report PIRS-663,1999.
  • 9Loftus P T, Weaver J T. Standardization of ~Co and 137 Cs Gamma-Ray Beams in Terms of Exposure [ J ]. J Res Nat Bureau Stand-A Phys and chem, 1974,78 ( A ) :465 - 476.
  • 10Rogers D W O, Bielajew A F, Nahum A E. Ion chamber response and Aw~l, correction factors in 6~Co beam by Monte Carlo simulation [ J ]. Phys Med Biol, 1985,30 (5) :429 -443.

引证文献1