期刊文献+

球囊保护左心室导线直接拔除法临床应用

Direct extraction of left ventricular lead with protection of balloon
原文传递
导出
摘要 目的观察经静脉左心室导线拔除的有效性及安全性。方法对4例拟拔除左心室导线的患者,应用在静脉造影球囊充盈的情况下,在X线透视下行左心室导线的直接拔除术。结果3例拔除导线手术完全成功;1例为临床成功;无并发症。结论球囊保护左心室导线直接拔除法是有效、安全的。 To observe efficacy and safety of transvenous left ventricular lead extraction. Method Using intervenous contrast filling balloon to apply direct left ventricular lead extraction directed by Xray on four patients needing left ventricular lead removal. Result Complete extraction was achieved in three cases, one case was clinical success, and no complication was observed. Conclusion Direct left vemricnlar lead extraction with nroteetion of balloon is effective and safe.
出处 《中华心律失常学杂志》 2012年第1期6-8,共3页 Chinese Journal of Cardiac Arrhythmias
关键词 心脏再同步治疗 左心室导线 导线拔除 球囊 Cardiac resynchronization therapy Left ventricular lead Lead extraction Balloon
  • 相关文献

参考文献16

  • 1mLinde C, Abraham WT, Gold MR, et al. Randomized trial of cardiac resynchronization in mildly symptomatic heart failure patients and in asymptomatic patients with left ventricular dysfunction and previous heart failure symptoms. J Am Coli Cardiol, 2008,52: 1834- 1843.
  • 2Wilkoff BL, Love CJ, Byrd CL, et al. Transvenous lead extraction: Heart Rhythm Society Expert Consensus on facilities, training, indications, and patient management. Heart Rhythm, 2009 , 6: 1085- 1104.
  • 3王方正,马坚,黄德嘉,黄从新,曹克将,华伟,沈法荣,吴立群,宿燕岗,杨杰孚,王景峰,陈泗林,陈柯萍,张澍.经静脉拔除心内膜导线:目前认识和建议(2011年修订版)[J].中华心律失常学杂志,2011,15(3):198-204. 被引量:29
  • 4Williams SE, Arujuna A, Whitaker J, et al. Percutaneous lead and system extraction in patients with cardiac resynchronization therapy ( CRT) devicesand coronary sinus leads. Pacing Clin Electrophysi- 01,2011,34: 1209-1216.
  • 5Kasravi B, Tobias S, Barnes MJ, et al. Coronary sinus lead extraction in the era of cardiac resynchronization therapy: single center experience. Pacing Clin Electrophysiol,2005 ,28 :51-53.
  • 6Martino GD, Orazi S, Bisignani G, et al. Safety and feasibility of coronary sinus left ventricular leads extraction: a preliminary report. J Int Card Electrophysiol, 2005 , 13 : 35 -38 ,.
  • 7AI - Khadm AS, Wilkoff BL. Extraction of transvenous pacemaker and defibrillator leads. In: Igor Sillger. Interventional Electrophysiolngy. Znd Edition. Philadelphia: Lippinentt Williams & Wilkins. 2001 :819-841.
  • 8Smith MC, Love CJ. Extraction of transvenous pacing and ICD lends. Pacing Clin Electrophysiol ,2008,31: 736-752.
  • 9Nerzil P , Taborsky M, Rezek Z, et al. Pacemaker and ICD lead extraction with elactrosurgical dissection sheaths and standard transvenous extraction systems: results of a randomized trial, Europace, 2007,9 :98-104.
  • 10Hamid S, Arujuna A, Khan S, et al. Extraction of chronic pacemaker and defibrillator leads from the coronary sinus: laser infrequendy used but required. Europace, 2009 , 11 : 213 -215.

二级参考文献40

  • 1唐闽,陈柯萍,王方正,华伟,张澍.29例起搏器和置入式心脏复律除颤器导线断裂的临床分析[J].中华心血管病杂志,2005,33(10):912-915. 被引量:17
  • 2Bilgutay AM, Jensen NK, Schmidt WR, et al. Incarceration of transvenous pacemaker electrode. Removal by traction. Am Heart J, 1969, 77:377-379.
  • 3Imparato AM, Kim GE. Electrode complications in patients with permanent cardiac pacemakers. Ten years' experience. Arch Surg, 1972, 105:705-710.
  • 4Madigan NP, Curtis JJ, Sanfelippo JF, et al. Difficulty of extrac- tion of chronically implanted tined ventricular endocardial leads.. J Am Coil Cardiol, 1984, 3:724-731.
  • 5Kratz JM, Leman R, Gillette PC. Forceps extraction of permanent pacing leads. Ann Thorac Surg, 1990, 49:676-677.
  • 6Byrd CL, Schwartz SJ, Sivina M, et al. Technique for the surgical extraction of permanent pacing leads and electrodes. J Thorac Car- diovasc Surg, 1985, 89:142-144.
  • 7Byrd CL, Schwartz SJ, Sivina M, et al. Experience with 127 pace- maker lead extractions. PACE, 1986, 9:282.
  • 8Byrd CL, Schwartz SJ, Hedin NB, et al. Intravascular lead extrac- tion using locking styles and sheaths. PACE, 1990, 13: 1871- 1875.
  • 9Fearnot NE, Smith HJ, Goode LB, et al. Intravascular lead extrac- tion using locking stylets, sheaths, and other techniques. PACE, 1990, 13:1864-1870.
  • 10A1-Khadra AS, Wilkoff BL. Extraction of transvenous pacemaker and defibrillator leads. In : Igor Singer. Interventional Electrophys- iology. 2nd Edition. Philadelphia:Lippincott Williams & Wilkins. 2001:819-841.

共引文献35