钙化积分扫描缩短冠脉CTA扫描范围及降低辐射量的作用 被引量:8

Effect of calcium scoring scan in reducing scan range and radiation dose in retrospectively ECG-gated CT coronary angiography
下载PDF
导出
摘要 目的:探讨使用钙化积分扫描缩短增强扫描范围以降低后门控冠状动脉CTA检查辐射量的临床应用价值。方法:40例患者行冠脉钙化积分及冠脉CTA检查,其增强扫描的实际扫描范围根据钙化积分扫描图像来决定,即冠状动脉树上缘1cm至心尖部心包影即将消失的层面。并按常规方法在定位像上测量增强扫描的预计扫描范围,即气管隆突下1cm至心脏下缘2cm。比较两种扫描范围及相应的辐射剂量的差异。结果:实际扫描范围比预计扫描范围平均缩短(2.09±0.76)cm(t=17.31,P<0.01),冠状动脉CTA各序列总剂量较常规方法平均减少(1.51±0.87)mSv,差异有极显著性意义(t=10.92,P<0.01),相当于各序列平均剂量总和的9.20%。结论:使用钙化积分扫描图像能有效缩短增强扫描的范围,从而明显降低冠状动脉CT血管成像的辐射量。 Objective:To find out a way of reducing scan range in retrospectively ECG-gated CT coronary angiography in order to reduce the radiation dose.Methods:Forty consecutive patients were analyzed retrospectively.The scan range of CT coronary angiography derived respectively from the scout view and from the axial images of calcium scoring were compared.The former was assumed to be determined from 1cm under the trachea carina to 2cm under the inferior margin of the heart.The latter was performed by identifying the left anterior descending coronary artery and posterior descending coronary artery and adding 1cm cranially and caudally.Effective radiation doses were calculated for CT coronary angiography using both scout view-derived and calcium scoring-derived scan range.Results:The calcium scoring-derived range was(10.40±0.60)cm;while the scout view-derived range was(12.5±0.94)cm which was significantly larger than that of using the calcium scoring-derived scan(t=17.31,P0.01).The average difference between them was(2.09±0.76)cm,corresponding to a radiation dose reduction of(1.67±0.72)mSv.After offsetting the dose given for the calcium scoring scan,the savings in radiation exposure when using a calcium scoring-derived instead of a scout view-derived scan range for CT coronary angiography-including the radiation dose of the calcium scoring scan and the test bolus scan-was(1.25±0.72)mSv(t=10.92,P0.01),which accounted for a dose reduction of 9.20%.Conclusion:Using the images from cal-cium scoring instead of the scout view to determine the scan range of contrast-unenhanced CT coronary angiography can effectively decrease the scan range and reduce the radiation dose significantly.
出处 《放射学实践》 2012年第3期305-308,共4页 Radiologic Practice
关键词 冠状动脉 钙化积分 扫描范围 辐射剂量 体层摄影术 X线计算机 回顾性心电门控 Coronary artery Calcium scoring Scan range Radiation dose Tomography X-ray computed Retrospectively electrocardiogram-gated
  • 相关文献

参考文献20

  • 1Bastarrika G,Lee YS,Huda W,et al.CT of coronary arterydisease[J].Radiology,2009,253(2):317-338.
  • 2Paul JF,Abada HT.Strategies for reduction of radiation dose incardiac multislice CT[J].Eur Radiol,2007,17(8):2028-2037.
  • 3张兆琪,徐磊.重视冠状动脉多层CT成像的低剂量检查[J].中华放射学杂志,2009,43(7):681-683. 被引量:80
  • 4Gopal A,Mao SS,Karlsberg D,et al.Radiation reduction withprospective ECG-triggering acquisition using 64-multidetectorcomputed tomographic angiography[J].Int J Cardiovasc Ima-ging,2009,25(4):405-416.
  • 5高建华,孙宪昶,戴汝平,王贵生,杨帆,韩文娟,李娜.64层容积CT冠状动脉成像技术及影响因素的分析[J].临床放射学杂志,2007,26(2):139-143. 被引量:16
  • 6Halpern EJ.Clinical cardiac CT:anatomy and function[M].NewYork:Thieme Medical Publishers,2008:16-19.
  • 7Zhang SZ,Hu XH,Zhang QW,et al.Evaluation of computedtomography coronary angiography in patients with a high heartrate using 16-slice spiral computed tomography with 0.37sgan-try rotation time[J].Eur Radiol,2005,15(6):1105-1109.
  • 8毛定飚,滑炎卿,王鸣鹏,张国桢,吴威岚,胡非,葛虓俊,丁其勇.心率及重建相位窗对多层螺旋CT冠状动脉造影图像质量的影响[J].中华放射学杂志,2004,38(5):521-524. 被引量:86
  • 9Schoepf UJ,Zwerner PL,Savino G,et al.Coronary CT angiogra-phy[J].Radiology,2007,244(1):48-63.
  • 10Hoffmann U,Ferencik M,Cury RC,et al.Coronary CT angiogra-phy[J].J Nuclear Medicine,2006,47(5):797-806.

二级参考文献48

  • 1高建华,孙宪昶,李剑颖,王贵生,李娜,戴汝平.后置滤过器C_2对64层螺旋CT冠状动脉成像质量的影响[J].中国医学影像技术,2007,23(1):73-76. 被引量:31
  • 2Husmann L,Valenta I,Gaemperli O,et al.Feasibility of lowdose coronary CT angiography:first experience with prospective ECG-gating Eur Heart J,2008,29:191-197.
  • 3Coles DR,Smail MA,Negus IS,et al.Comparison of radiation doses from multislice computed tomography coronary angiography and conventional diagnostic angiography.J Am Coll Cardiol,2006,47:1840-1845.
  • 4Brenner DJ,Hall EJ.Computed tomography:an increasing source of radiation exposure.N Engl J Med,2007,357:2277-2284.
  • 5Hausleiter J,Meyer T,Hermann F,et al.Estimated radiation dose associated with cardiac CT angiography.JAMA,2009,301:500-507.
  • 6Kalra MK,Maher MM,Toth TL,et al.Strategies for CT radiation dose optimization.Radiology,2004,230:619-628.
  • 7Mollet NR,Cademartiri F,de Feyter PJ.Non-invasive multislice CT coronary imaging.Heart,2005,91:401-407.
  • 8Szucs-Farkas Z,Kurmann L,Strautz T,et al.Patient exposure and image quality of low-dose pulmonary computed tomography angiography:comparison of 100-and 80-kVp protocols.Invest Radiol,2008,43:871-876.
  • 9Abada HT,Larehez C,Daoud B,et al.MDCT of the coronary arteries:feasibility of low-dose CT with ECG-pulsed tube current modulation to reduce radiation dose.AJR,2006,186:S387-S390.
  • 10Sehoenhagen P.Back to the future:coronary CT angiography using prospective ECG triggering.Eur Heart J,2008,29:153-154.

共引文献193

同被引文献86

  • 1赵飞,李磊,彭婉琳,敬茜,张凯,刘科伶,李玉明,胡斯娴,徐旭,李真林.呼气末屏气胸部CT扫描在体检中的应用价值[J].放射学实践,2020,0(2):238-241. 被引量:6
  • 2卢洋,王春丽.X线的辐射危害及防护安全管理[J].医学信息(医学与计算机应用),2016,29(30):26-27. 被引量:3
  • 3Furlow B. Radiation dose in computed tomography [J]. Ra diol Technol, 2010, 8(5):437- 450.
  • 4Shapiro BP, Young PM, Kamor B, et al. Radiation dose re- duction in CT coronary angiography [J]. Curr Cardiol Rep, 2010, 12(1): 59-67.
  • 5Henzler T, Fink C, Sehoenherg SO, et al. Dual-emergency CT: radiation dose aspects [J]. AJR, 2012, 199(5) : 16-25.
  • 6Becker CR. Advances in cardiac imaging [J]. Eur Radiol, 2003, 13(3): 50-52.
  • 7Nieman K. Can CT angiography replace catheter coronary an giography? [J]. Eurolntervention, 2010, 6: 65-71.
  • 8Cha KS, Kim MH, Kim YD. et al. Combined right transra dial coronary angiography and selective carotid angiography safety and feasibility in unselected patients[J].Catheter Car diovasc Interv, 2001, 53(3): 380-385.
  • 9Church TS, Levine BD, McGuire DK, et al. Coronary artery calcium score, risk factors, and incident coronary heart dis- ease events[J]. Atherosclerosis, 2007, 190(1) :224 ~231.
  • 10I-Ieyer CM,Mohr PS,Lemburg SP,et al.Image quality and radiation exposure at pulmonary CT angiography with 100-or 120-kVp protocol:prospective randomized study [J].Radiology,2007,245:577-583.

引证文献8

二级引证文献24

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部