摘要 目的评价迈瑞BC-5300全自动血细胞分析仪的性能。方法对BC-5300的精密度、携带污染率、线性、重复性进行测定;将100例健康体检标本同时用Sysmex XT-1800i和BC-5300测定白细胞(WBC)、红细胞(RBC)、血红蛋白(Hb)、血小板(PLT)、红细胞比容(HCT)5项指标并进行比对;并将这100例标本的BC-5300分类结果与手工分类结果进行比较。结果 BC-5300的精密度、携带污染率、线性、重复性均在允许范围内(仪器厂家提供);BC-5300和Sysmex XT-1800i对WBC、RBC、Hb、PLT、HCT的测定结果差异无统计学意义(P>0.05);BC-5300分类结果与手工分类结果进行比较,除单核细胞、嗜碱性粒细胞外,相关性较好;对异常细胞检出能力强。结论 BC-5300是一种较理想的全自动血细胞分析仪。 Objective To evaluate the function of the Mindray BC-5300 automated hematology analyzer.Methods Precision,carrier rate,linearity and repeatability were analyzed.Blood samples from 100 healthy people were tested by Sysmex XT-1800i and the BC-5300 respectively.Five indices of WBC,RBC,HGB,PLT and HCT were compared in two tests.The results differentiated by BC-5300 were compared with those by handwork for the screening of the abnormal cells.Results Precision,linearity,carrier rate and repetition studies of BC-5300 were all in the allowable ranges(supplied by factory authorities).The results of WBC,RBC,Hb,PLT and HCT had no significant difference between BC-5300 and Sysmex XT-1800i(P0.05).The classification results by BC-5300 were related to manual classification except Mon,Bas outside.In had good detection ability for abnormal cell.Conclusion The BC-5300 is a kind of ideal automatic blood cells analyzer.
作者 李彬 陈泽琴
出处 《检验医学与临床》 CAS 2012年第6期644-645,648,共3页 Laboratory Medicine and Clinic
关键词 BC-5300 精密度 携带污染率 线性 重复性 BC-5300 precision carrier rate linearity repeatability
  • 相关文献

参考文献6

  • 1International Council for Standardization in Hematology.Guidelines for the evaluation of blood cell analyzer inclu-ding those used for differential leukocyte and reticulocytecounting and cell marker application[J].Clin Lab Hema-tol,1994,2:157-174.
  • 2Waiters J,Patricia G.Performance evaluation of the Sys-mexXE-2100 hematology analyzer[J].Lab Hematol,2000,6:83-92.
  • 3Nakul AD,Sudaka SL,Ferrero VC,et al.Evaluation ofthe Sysmex Xe-2100hematology analyzer in hospital use[J].Clin Lab Anal,2003,17:113-123.
  • 4Lehto T,Hedberq P.Performance evaluation of AbbottCELL-DYN Ruby for routine use[J].Lab Hematol,2008,30:400-407.
  • 5Ruzicka K,Veitl M.The new hematology analyzer Sys-mex XE-2100[J].Archives of Pathology&LaboratoryMedicine,2001,125(3):391.
  • 6彭黎明,府伟灵,高行军.Sysmex SE-9000全自动血液分析仪性能的评价[J].临床检验杂志,2004,22(5):396-397. 被引量:10

二级参考文献8

  • 1International Council for Standardization in Haematology. Guidelines for the evaluation of blood cell analyzer including those used for differential leukocyte and reticulocyte counting and cell marker application. Prepared by the ICSH expert panel on cytometry [ J ].Clin Lab Haemat, 1994,16:157-174.
  • 2National Committee for Clinical Laboratory Standards. Leukocyte Differential Counting NCCLS Approved Standard H20-A2 [ S ].Villanova, PA, 1992.
  • 3Harrison P, Horton A, Grant D, et al. Immunoplatelet counting: a proposed new reference procedure [ J ]. Br J Haematol, 2000, 108:228-235.
  • 4Hallawell R,O Malley C,Hussein S,et al. An evaluation of the Sysmex NE-8000 hematology analyzer[J]. Am J Clin Pathol ,1991,96 :594-601.
  • 5Carter A B, Chavda N, Grant D,et al. Evaluation of the Sysmex SE-9000 automated hematology analyzer[ J]. Sysmex J Int, 1995,5:6-16.
  • 6Vives-Corrons J L,Besson I,Jou J M,et al. Evalution of the Abbott Cell-DYN 3500 hematology analyzer in a university hospital[ J]. Am J Clin Pathol, 1996,105:553-559.
  • 7Thalhammer-Seherrer R, Knobl P, Korninger L, et al. Automated fivepart white blood cell differential counts, efficiency of softwaregenerated white blood cell suspect flags of the hematology analyzers Sysmex SE-9000, Sysmex NE-8000, and Coulter STKS [ J ]. Arch Pathol Lab Med, 1997,121:573-577.
  • 8Bentley S A, Johnson A, Bishop C A. A parallel evaluation of four automated hematology analyzers [ J ]. Am J Clin Pathol, 1993,100:626-632.

共引文献9

同被引文献10

引证文献2

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部