中医脉象仪加压检测方法的研究初探 被引量:1

Study on the Pressure Methods of Pulse Detecting Instrument
下载PDF
导出
摘要 目的:探讨压力式脉象传感器正向加压法和反向加压法两种脉象检测方法对于测试结果的影响。方法:利用ZM-Ⅲc型脉象仪对于同一组受试者分别进行反向加压与正向加压两种方法进行操作测试,通过对两种不同方法下所加压力漂移值大小的对比,以及最佳取脉压力下各脉图参数的对比,比较两种方法的优劣。结果:正向加压法和反向加压法所产生的压力漂移值差异有统计学意义(P<0.05);两种方法最佳取脉压力下,各幅值参数除去h1外,其余P值均大于0.05,差异无统计学意义;脉象主峰波h1测量结果有影响(P<0.05);两种测量方法下脉搏波时间参数的变化情况差异无统计学意义(P>0.05)。结论:反向加压测试方法更稳定,有利于脉象测量结果的正确性。 Objective: To measure pressure which was taken to reverse the method and the positive pressure pulse method with the pressure pulse sensor for the comparative experiments for the test results. Methods: ZM-Ⅲc-type pulse instrument for the same group of subjects’ pulse information was tested by two ways, the positive and reverse pressure to operate. Under the two different methods increase the size of the contrast value of the pressure shift, and the best to take the pulse pressure pulse parameters of contrast, compare the pros and cons of both approaches. Results:The two methods of positive pressure and the reverse pressure generated by the pressure shift value P0.05, significant difference; two ways to best take the pulse pressure amplitude parameters to remove the h1, the remaining p-values are greater than 0.05, no significant difference; h1 pulse wave measured peak impact (P0.05); two measurements under the pulse wave changes in the parameters of time no significant difference (P 0.05). Conclusion: The reverse pressure test is more stable and conducive to the accuracy of pulse measurements.
机构地区 上海中医药大学
出处 《中国医学创新》 CAS 2012年第9期70-72,共3页 Medical Innovation of China
基金 上海浦东新区中医药事业发展专项资金项目(PZX2010-1) 上海市重点学科(第三期)中医诊断学经费(No.S30302) 上海中医药大学校级课题(L1161)
关键词 脉象仪 脉象传感器 脉象检测 加压方法 Pulse detecting system Pulse senor Method of pulse detecting pressurization pressure
  • 相关文献

参考文献4

二级参考文献6

共引文献34

同被引文献12

引证文献1

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部