摘要
建立了气相色谱法同时测定染发剂中邻位、间位、对位苯二胺的方法。 方法线性范围: 邻苯二胺0.3 ~40 .0 g/ L, 对苯二胺0.2 ~50 .0 g/ L, 间苯二胺0.4 ~70 .0 g/ L; 平均回收率: 苯二胺94 .7 % , 对苯二胺92 .2 % , 间苯二胺90 .1% 。 该法灵敏度高、操作简便、准确。
A method for simultaneous determination of (o,m,p _)penylenediamines in tint by gas chromatography was established. The result showed that the linear ranges of calibration curves were 0.3~40.0 g/L for o _ phenylenediamine, 0.2~50.0 g/L for p _ phenylenediamine, 0.4~70.0 g/L for m _ phenylenediamine; recoveries were 94.7%, 92.2% and 90.1% for o _ phenylenediamine, p _ phenylenediamine and m _ phenylenediamine, respectively. The method is simple, accurate and sensitive.
出处
《分析测试学报》
CAS
CSCD
2000年第1期79-80,共2页
Journal of Instrumental Analysis
关键词
苯二胺
气相色谱法
染发剂
Phenylenediamines
Gas chromatography
Tint