加成型硅橡胶一次法 两次法印模三维精度比较 被引量:8

Three-dimensional accuracy of one-step versus two-step putty-wash addition silicone impression technique
下载PDF
导出
摘要 目的比较加成型硅橡胶一次法、两次法印模的精度。方法采用重、轻体成分的加成型硅橡胶采用一次法和两次法分别制取同一实验模具的印模,检验比较所取印模对已行烤瓷全冠预备的21牙冠颈缘复制精度上的差异;灌注石膏模型后,游标卡尺测量主模型及石膏模型上各标志点的距离,对测量数据进行统计学分析。结果两次法印模在对牙冠颈缘复制精度上效果优于一次法印模,两者有显著差异。在线性变化率上比较,一次法印模的变化率均比两次法要大,但与原始数据比较无明显差异。结论加成型硅橡胶印模材料使用两次法印模优于一次法印模。 Objective To compare the three-dimensional accuracy of one-step putty-wash with two-step putty-wash addition silicone impression techniques. Methods A crystal master model containing one full ceramic crown abutment preparation(at no.21 position) was used as the positive control.The impressions of the master model were made using one-step and two-step putty-wash technique,respectively.Rank-sum test was employed to evaluate the variance between the control model and its impressions made by two techniques.The stones dies were poured from impressions of the master model.Vertical and horizontal dimensions between marks on the master model and stone dies were measured with electronic caliper.Data were statistically analyzed to compare the means of percentage changes.Results The accuracy of replication in shoulder continuity of two-step impressions was higher than that of one-step impressions,and the difference was statistically significant.The percentage deviation of one-step technique was bigger than that of two-step technique;however there was no statistical significance.Conclusion Three-dimensional accuracy of the two-step putty-wash addition silicone impression technique is higher than one-step impression technique.
出处 《安徽医学》 2012年第3期271-273,共3页 Anhui Medical Journal
关键词 加成型硅橡胶 重/轻体 三维精度 Addition silicone Putty-wash Three-dimensional accuracy
  • 相关文献

参考文献9

  • 1Levartovsky S,Folkman M,Alter E. Elastomeric impression materials[J].Refuat Hapeh Vehashinayim(Israel Journal of Dental Medicine),2011,(02):54-64,75.
  • 2Hasson AK. Dimensional accuracy of 3 silicone dental impression materials[J].Eastern Mediterranean Health Journal,2006,(05):632-636.
  • 3杜方翀,陈吉华,甘云娜,林玲,成炜,沈焕.五种硅橡胶印模材料对牙冠颈缘复制精度的比较[J].中国美容医学,2007,16(4):501-503. 被引量:11
  • 4文理衡,刘清辉,周琳.固定修复三种印模法临床效果的对比[J].现代口腔医学杂志,2005,19(2):116-116. 被引量:11
  • 5杜方翀,陈吉华,甘云娜.戊二醛消毒液对4种高分子印模尺寸变化的影响[J].实用口腔医学杂志,2007,23(2):268-271. 被引量:11
  • 6Federick DR,Caputo A. Comparing the accuracy removable hydrocolloid and elastomeric impression materials[J].Journal of the American Dental Association,1997,(02):183-188.
  • 7Richards MW,Zeiaei S,Bagby MD. Woking times and dimensional accuracy of the one-step putty/wash impression technique[J].J Prosthodon,1998,(04):250-255.
  • 8Caputi S,Varvara G. Dimensional accuracy of resultant casts made by a monophase,one-step and two-step,and a novel two-step putty/light-body impression technique:an in vitro study[J].Journal of Prosthetic Dentistry,2008,(04):274-281.
  • 9American National Standards/American Dental Association Specifications. Council on Dental Materials,Instruments,and Equipment[J].Journal of the American Dental Association,1983,(04):640-641.

二级参考文献13

  • 1文理衡,刘清辉,周琳.固定修复三种印模法临床效果的对比[J].现代口腔医学杂志,2005,19(2):116-116. 被引量:11
  • 2Livaditis Gus J.The matrix impression system for rixed prothodonties.J Prosthet Dent,1998,79:208-216.
  • 3Federick DR,Caputo A. Comparing the accuracy removable hydrocolloid and elastomaric impression materials[J].J Am Dent Assoc,1997,128(2): 183-188.
  • 4Richards MW, Zeiaei S, Bagby MD, et al. Working times and dimensional accuracy of the one-step putty wash impression technique[J]. J Prosthodont,1998,7(4):250-254.
  • 5Di Felice R, Scotti R.The influence of the mixing technique onthe content of voids in two polyether impression materials[J]. Schweiz M onatsschr Zahnmed,2002,112(1): 12-16.
  • 6Livaditis Gus J.The matrix impression system for fixed prothodonties[J].J Prosthet Dent,1998,79:208-216.
  • 7Craig RG. Restorative dental materials [M]. 9th ed.St Louis:Mosby,1993,142-143.
  • 8Femandes CP,Vassilakos N. Accuracy, detail reproduction, and hardness of gypsum casts produced from silicone impressions treated with glow discharge[J]. J Prosthet Dent,1993, 70(5):457-464.
  • 9Tjan AH,Nemetz H,Nguyen LT,et al.Effect of tray space on the accuracy of monophasic polyvinyl silocane impressions.J Prosthet Dent,1992,68(1):19
  • 10Gordon GE,Johnson GH,Drennon DG.The effect of tray selection on the accuracy of elastomeric impression materials.J Prosthet Dent,1990,63(1):12

共引文献28

同被引文献513

引证文献8

二级引证文献28

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部