期刊文献+

对用增量法检测尿微量白蛋白测定试剂盒线性范围的探讨 被引量:2

Discussion of the microalbuminuria determination kits linear Range Detection with increment methodds
下载PDF
导出
摘要 目的通过对尿微量白蛋白测定试剂盒(乳胶比浊法)线性范围的测定,简介试验时能否用增量法(样品量加倍模式)达到线性高点。方法用固定时间法在迈瑞BS-800全自动生化分析仪上对线性范围进行检测。结果该试剂盒不能通过增量法达到线性高点。结论应选择浓度适当的样品进行试剂(盒)线性范围试验。 Objective Through the determination of microalbuminuria kits(latex turbidity method) linear range,to introduce if the linear high can be reached with incremental method(sample amount double mode) during the experiment.Methods To test the linear range with the fixed time method with the Mindray BS-800 automatic biochemistry analyzer.Results The kit could not reach the linear high with incremental method.Conclusion Appropriate concentration samples should be selected to conduct the reagent(box) linear range test.
作者 张蓓 何乐春
出处 《中国医药导报》 CAS 2012年第12期118-119,共2页 China Medical Herald
关键词 线性范围 增量法 前后带现象 Linear range Incremental method Before and after phenomenon
  • 相关文献

参考文献8

二级参考文献13

共引文献212

同被引文献26

  • 1中华人民共和国卫生部.三级综舍医院评审标准.北京:人民卫生出版社,2011.
  • 2Folsom AR, Catellier D. C-reactive protein apd incidentcoronoryheart disease in the atherosclerosis risk in coniniy-pities study,Am Heart J,2002,144(2) ;233.
  • 3Brady GR,Fienup JR. Measurement range of phase retrieval inop-tical inoptical suiface andwavefront metrology. Appl Opt,2009,48(3);442449.
  • 4Rifai N, Ridker PM. High-sensitivity C-reactive protein: a novelsnd promising marker of coronary hemx disease. Clin Chem,2001,47(3) :403411.
  • 5Pearsin TA,Mensah GA, Alexander RW,et al. Markers of inflam-mation and cardiovascular disease : applicationto clinical and pub-lic health practice: a statement for healthcare professionals fromthe centers for disease control and preventionand the americanheart association. Circulation,2003 ,107(3) :499-511.
  • 6Laverdiere M,Restieri C,Habel F. Evaluationof the in vitro activi-ty of caspofungin against bloodstream isolates of Candida speciesfromcancer patients: comparison of Etest and NCCLS referencemethods. Int J Antimicrob Agents,2002,20 ( 6) ;468-471.
  • 7Clinical and Laboratory Standards Institute. EP6-A2,Evaluation of the Linearity of Quantitative Measurement Procedures:a statistical approach[S].Wayne PA:CLSI,2003.doi:10.1016/j.jbiotec.2009.10.004.
  • 8冯仁丰.临床检验质量管理技术基础[M]上海:上海科学技术文献出版社,200739-50.
  • 9Clinical and Laboratory Standards Institute. EP9-A2,Method Comparison and Bias Estimation Using Patient Samples:a statistical approach[S].Wayne PA:CLSI,2002.doi:10.1148/rg.301095110.
  • 10Kroll MH,Praestgaard J,Michaliszyn E. Evaluation of the extent of nonlinearity in reportable range studies[J].Archives of Pathology and Laboratory Medicine,2000,(09):1331-1338.