摘要
在惯性约束聚变(ICF)实验研究中,科学级CCD常和光谱仪、条纹相机等诊断设备组合进行冲击波等物理过程的时间、空间方面的诊断研究。非线性作为CCD一个非常重要的性能参数,在实验的后处理过程中有着相当重要的作用,实验分别采取固定曝光时间,改变光源的出射功率密度、固定光源的输出功率密度和改变CCD的曝光时间两种方案对TEK 512×512DB CCD的非线性特性进行测试,计算得到两种实验方案下该CCD的非线性度分别为0.595%和0.508%,并通过对各种因素的评估得到第一种标定方案为最佳,为科学级CCD的性能参数自动测试平台的建立提供参考。
In inertial confinement fusion(ICF) experimental studies,scientific-grade CCD coupled with streak camera is widely used in time and space diagnosis experiment with physical processes such as shock wave.As a significant property of the scientific grade CCD,nonlinearity plays an important role in experimental dealing.The experiment tests the nonlinearity property of the TEK 512×512DB CCD with the two schemes respectively: that the exposure time is fixed,the power density of the RS-5B is varied;power density is fixed,the exposure time is varied,and it shows the nonlinearity of the CCD is 0.595% and 0.508% respectively,meanwhile,the two experimental scheme are evaluated by analyzing kinds of factors,indicating that the first scheme is better,which provides the reference for establishment of automated testing platform.
出处
《光学学报》
EI
CAS
CSCD
北大核心
2012年第4期39-44,共6页
Acta Optica Sinica
基金
国家自然科学基金(11005097)
中国工程物理研究院科学技术发展基金(2010A0102003)
中南大学研究生教育创新工程基金(2010ssxt128)资助课题
关键词
测量
科学级CCD
条纹相机
非线性
measurement
scientific-grade CCD
streak camera
nonlinearity