摘要
目的:研究钛表面电弧离子镀膜的结合强度。方法:以德国工程师手册(VDI3198)为标准,通过压入实验测量膜结合强度,利用扫描电镜观察膜结构及其与基体的结合情况。结果:载荷为400N时,膜的断裂模式属于HF3;扫描电镜可见膜与基体结合良好、厚度均匀、平整,膜厚度约为1.2μm。结论:钛表面电弧离子镀膜与基材间的结合强度能够满足临床需要。
PURPOSE: To study the bonding strength between the FCVA(filtered cathode vacuum arc) film and the titanium substrate.METHODS: According to the Engineer Manual of Germany(VDI3198),the indentation test was applied to evaluate the bonding strength between the FCVA film and the titanium substrate.The structure of the FCVA film and the interface between the film and the substrate were observed by means of scanning electron microscope(SEM).RESULTS: The fracture pattern of the film was classified as HF3 at the load of 400N,the bonding strength was enough between the FCVA film and the titanium substrate.SEM photograph showed the thickness of the FCVA film was about 1.2μm equably and the film bonded to the substrate well.CONCLSION: The bonding strength between the FCVA film and the titanium substrate was enough to meet clinical demand.
出处
《上海口腔医学》
CAS
CSCD
北大核心
2012年第2期127-129,共3页
Shanghai Journal of Stomatology
基金
国家自然科学基金(30171022)
陕西省科技攻关课题(2003K10G27)~~
关键词
电弧离子镀
钛
表面处理
结合强度
Filtered cathode vacuum arc
Titanium
Surface treatment
Bonding strength