摘要 在医用X-CT中,散射现象对重建图像的质量有重要的影响。多源锥束CT是医用CT的一种发展趋势,可成倍提高时间分辨率。但与单源CT相比,多源CT额外的光源会导致散射现象的加重,因此在实际应用中需要着重考虑这一问题。本文在开源工具EGSnrc的基础上,开发了一套基于蒙特卡罗法的CT仿真软件SimProj,用该软件获得了贴近实际的三源锥束CT投影数据。重建图像显示出了预期的散射伪影,结果分析对优化三源CT的设计和制造有重要的指导意义。 The scattering has serious influence on the image quality in X-ray CT.Multi-source CT is a trend of future CT system since it can significantly improve the temporal resolution.However,as the number of sources increases,more scattered radiation will occur.The scattering issue should be taken into consideration in the practical applications.A CT simulation software,SimProj,has been developed based on the open source tool EGSnrc to generate cone-beam projections which approach the real data.Reconstructed images from the simulated data show scattering induced artifacts as expected.The analysis results are important and helpful for the optimization of designing and manufacturing of triple-source cone-beam CT.
出处 《CT理论与应用研究(中英文)》 2012年第1期1-9,共9页 Computerized Tomography Theory and Applications
基金 973项目(2010CB834302) 国家自然科学基金(30570511 30770589)
关键词 蒙特卡罗 X线 三源锥束CT 散射 EGSNRC Monte Carlo X-ray triple-source cone-beam CT scattering EGSnrc
  • 相关文献

参考文献17

  • 1Glover GH.Compton scatter effects in CT reconstructions[J].Medical Physics,1982,9(6):860-867.
  • 2Joseph PM,Spital RD.The effects of scatter in x-ray computed tomography[J].Medical Physics,1982,9:464.
  • 3Siewerdsen JH,Jaffray DA.Cone-beam computed tomography with a flat-panel imager:Magnitudeand effects of x-ray scatter[J].Medical Physics,2001,28(2):220-231.
  • 4Endo M,Tsunoo T,Nakamori N,et al.Effect of scattered radiation on image noise in conebeam CT[J].Medical Physics,2001,28(4):469-474..
  • 5唐杰,张丽,高文焕.基于平板探测器的锥束CT系统综述[J].中国体视学与图像分析,2004,9(2):65-70. 被引量:19
  • 6Flohr TG,McCollough CH,Bruder H,et al.First performance evaluation of a dual-source CT(DSCT)system[J].European Radiology,2006,16(2):256-268.
  • 7Zhao J,Jin YN,Lu Y,et al.A filtered backprojection algorithm for triple-source helicalcone-beam CT[J].Ieee Transactions on Medical Imaging,2009,28(3):384-393.
  • 8Bruder H,Stierstorfer K,Raupach R,et al.Design considerations in cardiac CT[C] //Proceedings of Medical Imaging 2006,San Deigo,CA,USA.
  • 9Engel KJ,Herrmann C,Zeitler G.X-ray scattering in single-and dual-source CT[J].MedicalPhysics,2008,35(1):318-332.
  • 10Kawrakow I,Rogers D.The EGSnrc code system:Monte Carlo simulation of electron and photontransport[R].NRC Institute for National Measurement Standards,2010.

二级参考文献21

  • 1曾凯,陈志强.三维锥形束CT解析重建算法发展综述[J].中国体视学与图像分析,2003,8(2):124-128. 被引量:12
  • 2汤少杰,牟轩沁,闫浩.基于医学X射线成像物理模型的模体投影仿真计算[J].西安交通大学学报,2006,40(8):901-905. 被引量:7
  • 3叶侠娟.[D].西北大学,.
  • 4许淑燕.蒙特卡罗方法在实验核物理中的应用[M].北京:原子能出版社,1996..
  • 5G.T.Herman.Image reconstruction from projections.Academic Press,INC.,1980.
  • 6Chye Hwang Yan,Robert T.Whalen,Gary S.Beaupre,Shin Y.Yen,Sandy Napel.Modeling of polychromatic attenuation using computed tomography reconstructed images.Med Phys.1999,26(4):631-642.
  • 7H.H.Barrette and W.Swindell Radiological Imaging(the theory of image formation.detection and processing vol.2,Academic Press.INC.,1981.
  • 8Feldkamp A, Davis L C and Kress J W. Praactical cone-beam algorithm[J].Opt Soe Am A, 1984,VoL 1,612-619.
  • 9Grangeat E Mathematical framework of cone beam 3-D reconstruction via the first derivative of the radon transform[C]. In Herman GT, Louis A K and Natterer F, Eds. Mathematical Methods in tomography. Springer verlag, 1990.
  • 10Smith B D. Image reconstruction from cone-beam projection: necessary and sufificient condition and reconstruction methods[J]. IEEE Trans Med Imag 1985, MI-4: 14-28.

共引文献39

同被引文献1

引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部