摘要
本法利用波长色散X射线荧光光谱仪测定灯用煤油中的硫 ,测量精密度较好 ,加标回收率为 90 .4 %~ 10 8.9% ,检测限 (LLD)为 0 .6× 10 - 6 (ws)。
Trace sulfur in illuminating kerosene was measured by sequential X-ray fluorescence spectrometer. The experimental results show that the precision are acceptable, and recovery is 90.4%~108.9%,the lower limit of detection (LLD) is 0.6ppm.
出处
《石油与天然气化工》
CAS
CSCD
北大核心
2000年第2期94-95,共2页
Chemical engineering of oil & gas