胶片扫描法对低能X射线标准辐射场的研究

Studay on Reference Radiation of Low-energy x-ray With Film Scanning
下载PDF
导出
摘要 采用胶片剂量计测量了低能X射线辐射装置距X射线光管焦斑不同距离辐射野的大小,得出在距离光管焦斑1 000 mm的参考面辐射野大小及均匀性,并与二维矩阵电离室扫描结果进行了对比。实验结果表明,在距X射线光源焦斑1 000 mm处,在保证直径为100 mm的均匀野的条件下,非均匀性不大于1.5%。在保证非均匀性不大于5%和50%条件下,均匀野大小分别为130 mm和145 mm,满足了低能X射线空气比释动能基准装置量值复现,国际比对和向下量值传递的要求。 The size of reference radiation field of low - energy x - ray was measured by film in different position from the tube spot, the diameter of the radiation field and its were obtained in reference plane ( 1 000 mm from the radiation source) and were compared with the results through two - dimensional ion chamber. The results show that the diameter of the radiation field is 100 mm within 1.5% ( 1 000 mm from the radiation source). The diamerers of theradiation field are 130 nun and 145 mm within 5% and 50% . These results well meet the requirements of the value repetition of the low - energy X - ray air kerma value standard device, international intereomparison and transfer of quantity values.
出处 《核电子学与探测技术》 CAS CSCD 北大核心 2012年第2期207-209,217,共4页 Nuclear Electronics & Detection Technology
关键词 X射线 胶片 标准辐射场 均匀性 X - ray, film, reference radiation, uniformity
  • 相关文献

参考文献4

  • 1IAEA. Radiation oncology physics: a handbook for teachers and students [ Z ]. Vienna : IAEA,2005.
  • 2Pai U, Indra J. Das TG -69: radiographic firm for megavoltage beam dosimetry[ J ]. Med Phys, 2007,34 (6) :2228 - 2257.
  • 3Chair AN, Blackwell CR, Coursey BM, et al. Radiochromic film dosimetry [ J ]. Med Phys, 1998, 25 (11) :2093-2115.
  • 4俞顺飞,程金生,李开宝,苏旭.放射治疗辐射场剂量分布胶片测量技术的研究进展[J].中华放射医学与防护杂志,2008,28(3):313-314. 被引量:3

二级参考文献17

  • 1马金利,傅小龙,蒋国梁.放射摄片用胶片剂量仪在肿瘤放射物理学中的应用[J].中华放射医学与防护杂志,2004,24(5):483-486. 被引量:7
  • 2IAEA. Radiation oncology physics: a handbook for teachers and students. Vienna: IAEA, 2005.
  • 3Pai U, Indra J. Das TG-69: radiographic film for megavohage beam dosimetry. Med Phys, 2007, 34(6) :2228-2257.
  • 4Chair AN, Blackwell CR, Coursey BM, et al. Radiochromic film dosimetry. Med Phys, 1998,25 ( 11 ) :2093-2115.
  • 5Ju SG, Ahn YC, Huh SJ, et al. Film dosimetry for intensity modulated radiation therapy: Dosimetric evaluation. Med Phys, 2002, 29(3) :351-355.
  • 6Bucciolini M, Bounamici FB, Casati M. Verification of IMRT fields by film dosimetry. Med Phys, 2004, 31(1):161-168.
  • 7Fiandra C, Ricardi U, Ragona R. Clinical use of EBT model Gafchromic^TM film in radiotherapy. Med Phys, 2006, 33(11):4314- 4319.
  • 8Cheung T, Butson MJ, Yu PK. Post-irradiation coloration of Gafchromic EBT radiochromic film. Phys Med Biol, 2005, 50(20): 281-285.
  • 9ISP, GAFCHROMIC EBT: SeLf-developing film for radiotherapy dosimetry. New Jersey:ISP,2005.
  • 10Danciu C, Proimus BS, Rosenwald JC, et al. Variation of sensitometfic curves of radiographic films in high energy photon beams. Med Phys, 2001, 28(6) :966-974.

共引文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部